Hand Therapy Clinic, Neurosciences and Trauma Program, Hamilton Health Sciences, Hamilton, Ontario, Canada.
J Hand Ther. 2012 Oct-Dec;25(4):358-61; quiz 362. doi: 10.1016/j.jht.2012.06.003. Epub 2012 Sep 10.
Clinical measurement study.
Skin temperature asymmetries (STAs) are used in the diagnosis of complex regional pain syndrome (CRPS), but little evidence exists for reliability of the equipment and methods.
This study examined the reliability of an inexpensive infrared (IR) thermometer and measurement points in the hand for the study of STA.
ST was measured three times at five points on both hands with an IR thermometer by two raters in 20 volunteers (12 normals and 8 CRPS).
ST measurement results using IR thermometers support inter-rater reliability: intraclass correlation coefficient (ICC) estimate for single measures 0.80; all ST measurement points were also highly reliable (ICC single measures, 0.83-0.91).
The equipment demonstrated excellent reliability, with little difference in the reliability of the five measurement sites. These preliminary findings support their use in future CRPS research.
Not applicable.
临床测量研究。
皮肤温度不对称(STA)用于复杂区域疼痛综合征(CRPS)的诊断,但对于设备和方法的可靠性证据有限。
本研究旨在检验一种廉价的红外(IR)温度计和手部测量点在手部 STA 研究中的可靠性。
20 名志愿者(12 名正常人和 8 名 CRPS 患者)由两名评估者在双手的五个部位使用 IR 温度计进行三次 STA 测量。
IR 温度计的 STA 测量结果支持评估者之间的可靠性:单次测量的组内相关系数(ICC)估计值为 0.80;所有 STA 测量点也非常可靠(ICC 单次测量值为 0.83-0.91)。
该设备表现出极佳的可靠性,五个测量部位的可靠性差异不大。这些初步发现支持它们在未来的 CRPS 研究中的应用。
不适用。