使用两种电子根尖定位仪对模拟根尖穿孔进行体外检测。

In vitro detection of simulated apical root perforation with two electronic apex locators.

作者信息

Nazari Moghaddam Kiumars, Nazari Shahrzad, Shakeri Leila, Honardar Kiamars, Mirmotalebi Farshid

机构信息

Department of Endodontics, Dental School, Shahed University of Medical Sciences, Tehran, Iran.

出版信息

Iran Endod J. 2010 Winter;5(1):23-6. Epub 2010 Feb 20.

Abstract

INTRODUCTION

The aim of this study was to evaluate the accuracy of two electronic apex locators (Smarpex and NovApex) in detecting apical perforation.

MATERIALS AND METHODS

After access cavity preparations, the working length was determined by the first examiner before and after perforation with a stereomicroscope by introducing a K-file size 10 into the canal and size 25 K-file up to the perforation site, respectively. The specimens were embedded in a 17-well plastic box containing alginate. The root canals were irrigated with chlorhexidine 0.2% (CHX) through a 27-gauge needle. Two examiners measured the root canal length twice and the mean value was calculated. The data were analyzed using Kolmogorov-Smirnov and ANOVA tests.

RESULTS

The results obtained with each Electronic Apex Locator were compared with the corresponding control length. The statistical analysis showed reliable accuracies in detecting the perforation site for the two experimental electronic apex locators (60% and 80% for NovApex and Smarpex, respectively).

CONCLUSION

Although no significant difference was shown between Smarpex and NovApex, these results suggest that electronic apex locators can effectively and reproducibly detect root canal perforations.

摘要

引言

本研究的目的是评估两种电子根尖定位仪(Smarpex和NovApex)检测根尖穿孔的准确性。

材料与方法

开髓后,由第一位检查者分别在穿孔前后通过立体显微镜确定工作长度,穿孔前将10号K锉插入根管,穿孔后将25号K锉插入至穿孔部位。标本嵌入含有藻酸盐的17孔塑料盒中。通过27号针头用0.2%洗必泰(CHX)冲洗根管。两位检查者测量根管长度两次并计算平均值。数据采用Kolmogorov-Smirnov检验和方差分析进行分析。

结果

将每种电子根尖定位仪获得的结果与相应的对照长度进行比较。统计分析表明,两种实验性电子根尖定位仪在检测穿孔部位方面具有可靠的准确性(NovApex和Smarpex分别为60%和80%)。

结论

虽然Smarpex和NovApex之间未显示出显著差异,但这些结果表明电子根尖定位仪能够有效且可重复地检测根管穿孔。

相似文献

4
Accuracy of Working Length Determination using NovApex and Root-ZX Apex Locators: An in vitro Study.
J Contemp Dent Pract. 2017 May 1;18(5):383-385. doi: 10.5005/jp-journals-10024-2051.
8
Comparative Evaluation of Different Apex Locators as a Diagnostic Tool to Detect Root Perforations: An Study.
J Pharm Bioallied Sci. 2024 Feb;16(Suppl 1):S828-S830. doi: 10.4103/jpbs.jpbs_1045_23. Epub 2024 Feb 29.
9
Accuracy and repeatability of 3 apex locators in locating root canal perforations: an ex vivo study.
J Endod. 2014 Aug;40(8):1241-4. doi: 10.1016/j.joen.2014.02.004. Epub 2014 Apr 21.

引用本文的文献

1
Factors influencing the accuracy of electronic apex locators: A scoping review.
Dent Res J (Isfahan). 2025 Jun 27;22:26. doi: 10.4103/drj.drj_634_23. eCollection 2025.
5
Influence of Root Canal Curvature on the Accuracy of Root ZX Electronic Foramen Locator: An Study.
Iran Endod J. 2017 Spring;12(2):173-178. doi: 10.22037/iej.2017.34.

本文引用的文献

1
The evaluation of four electronic apex locators in teeth with simulated horizontal oblique root fractures.
J Endod. 2008 Dec;34(12):1497-9. doi: 10.1016/j.joen.2008.09.002. Epub 2008 Oct 18.
2
Mineral trioxide aggregate as repair material for furcal perforation: case series.
J Endod. 2008 Sep;34(9):1130-3. doi: 10.1016/j.joen.2008.05.019. Epub 2008 Jul 14.
3
Comparison of radiographic and electronical working length determination on palatal and mesio-buccal root canals of extracted upper molars.
Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008 Aug;106(2):e90-3. doi: 10.1016/j.tripleo.2008.04.007. Epub 2008 Jun 13.
4
Early flaring and crown-down shaping influences the first file bind to the canal apical third.
Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008 Aug;106(2):e99-101. doi: 10.1016/j.tripleo.2008.04.010. Epub 2008 Jun 13.
5
Ex vivo evaluation of the accuracy of two electronic apex locators during root canal length determination in primary teeth.
Int Endod J. 2008 Apr;41(4):317-21. doi: 10.1111/j.1365-2591.2007.01366.x. Epub 2008 Jan 19.
6
Accuracy of two different apex locators in primary teeth with and without root resorption.
Clin Oral Investig. 2008 Jun;12(2):137-41. doi: 10.1007/s00784-007-0157-5. Epub 2007 Nov 20.
7
An in vitro evaluation of the accuracy of Dentaport ZX apex locator in enlarged root canals.
Aust Dent J. 2007 Sep;52(3):193-7. doi: 10.1111/j.1834-7819.2007.tb00488.x.
8
Influence of apical constriction diameter on Root ZX apex locator precision.
J Endod. 2007 Aug;33(8):995-8. doi: 10.1016/j.joen.2007.05.009.
9
In vivo evaluation of ProPex electronic apex locator.
J Endod. 2007 Aug;33(8):974-7. doi: 10.1016/j.joen.2007.04.003. Epub 2007 Jun 22.
10
Evaluation of precision of length determination with 3 electronic apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator, and RomiAPEX D-30.
Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2007 Oct;104(4):e91-4. doi: 10.1016/j.tripleo.2007.05.016. Epub 2007 Aug 20.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验