离体条件下使用简易安装模型评估三种电子根尖定位仪的准确性和重复性系数:初步报告。

Ex vivo evaluation of the accuracy and coefficient of repeatability of three electronic apex locators using a simple mounting model: a preliminary report.

机构信息

Department of Operative Dentistry, Federal University of Paraíba, João Pessoa, Paraíba, Brazil.

出版信息

Int Endod J. 2010 Apr;43(4):269-74. doi: 10.1111/j.1365-2591.2009.01674.x.

Abstract

AIM

To compare ex vivo the accuracy and coefficient of repeatability of three electronic apex locators in locating the apical constriction.

METHODOLOGY

Thirty-one single-rooted teeth were used. The teeth were sectioned at the cement--enamel junction. A mounting model was used for the measurement of electronic length (EL). The Root ZX-II, the Mini Apex Locator and the Novapex were used for electronic measurements. Each electronic measurement was obtained and repeated. After the last measurement, the file was cemented in place, and the apical 4 mm of each root canal was exposed. The distance from the tip of the file to the apical constriction was determined by three investigators and compared with the corresponding ELs.

RESULTS

The coefficient of repeatability of all devices was acceptable: Root ZX-II, 0.04 mm; Mini Apex Locator, 0.10 mm; and Novapex, 0.08 mm. There was little variation in inter-examiner agreement; the rho(C) (Lin) correlation coefficient was 0.83 for examiners 1 and 2, 0.88 for examiners 1 and 3 and 0.99 for examiners 2 and 3. Using the Root ZX-II, 13 of 31 electronic measurements were located at the apical constriction (42%). Otherwise, the tip of the file was not located at the apical constriction in any of the electronic measurements with the other two devices. The Wilcoxon signed rank test did not reveal any statistical difference between the Root ZX-II measurements and the actual length (P = 0.628), but there was a statistical difference between the Mini Apex Locator and Novapex measurements and the actual length position (P < 0.05).

CONCLUSIONS

The devices tested in this study had a high coefficient of repeatability. The Root ZX-II was accurate, but the Mini Apex Locator and Novapex were not accurate in locating the apical constriction.

摘要

目的

比较三种电子根尖定位仪在定位根尖狭窄部时的准确性和可重复性系数。

方法

使用 31 颗单根牙。将牙齿在牙釉质-牙骨质交界处切割。使用安装模型进行电子长度(EL)测量。使用 Root ZX-II、Mini Apex Locator 和 Novapex 进行电子测量。每次测量后,将器械粘固在原位,暴露每个根管的根尖 4mm。由三位研究者确定距器械尖端至根尖狭窄部的距离,并与相应的 EL 进行比较。

结果

所有设备的可重复性系数均在可接受范围内:Root ZX-II,0.04mm;Mini Apex Locator,0.10mm;Novapex,0.08mm。三位研究者之间的一致性变化很小;Lin 相关系数(rho(C))在 1 号和 2 号检查者之间为 0.83,在 1 号和 3 号检查者之间为 0.88,在 2 号和 3 号检查者之间为 0.99。使用 Root ZX-II,31 次电子测量中有 13 次(42%)定位在根尖狭窄部。在其他两种设备的任何电子测量中,器械尖端都未定位在根尖狭窄部。Wilcoxon 符号秩检验未发现 Root ZX-II 测量值与实际长度之间存在统计学差异(P=0.628),但 Mini Apex Locator 和 Novapex 测量值与实际长度位置之间存在统计学差异(P<0.05)。

结论

本研究中测试的设备具有较高的可重复性系数。Root ZX-II 是准确的,但 Mini Apex Locator 和 Novapex 定位根尖狭窄部不准确。