Department of Chemistry, University of Warwick, Coventry, United Kingdom.
Anal Chem. 2013 Jan 2;85(1):163-9. doi: 10.1021/ac3023586. Epub 2012 Dec 11.
The electrochemical measurement of dopamine (DA), in phosphate buffer solution (pH 7.4), with a limit of detection (LOD) of ∼5 pM in 50 μL (∼ 250 attomol) is achieved using a band electrode comprised of a sparse network of pristine single-walled carbon nanotubes (SWNTs), which covers <1% of the insulating substrate. The SWNT electrodes are deployed as amperometric (anodic) detectors in microfluidic cells, produced by microstereolithography, designed specifically for flow injection analysis (FIA). The flow cells, have a channel (duct) geometry, with cell height of 25 μm, and are shown to be hydrodynamically well-defined, with laminar Poiseuille flow. In the arrangement where solution continuously flows over the electrode but the electrode is only exposed to the analyte for short periods of time, the SWNT electrodes do not foul and can be used repeatedly for many months. The LOD for dopamine (DA), reported herein, is significantly lower than previous reports using FIA-electrochemical detection. Furthermore, the SWNT electrodes can be used as grown, i.e., they do not require chemical modification or cleanup. The extremely low background signals of the SWNT electrodes, as a consequence of the sparse surface coverage and the low intrinsic capacitance of the SWNTs, means that no signal processing is required to measure the low currents for DA oxidation at trace levels. DA detection in artificial cerebral fluid is also possible with a LOD of ∼50 pM in 50 μL (∼2.5 fmol).
在 pH 值为 7.4 的磷酸盐缓冲溶液中,使用由稀疏的原始单壁碳纳米管 (SWNT) 网络组成的带电极对多巴胺 (DA) 进行电化学测量,在 50μL(约 250 飞摩尔)中检测限 (LOD) 约为 5 pM。该 SWNT 电极覆盖 <1% 的绝缘基底,用作微流控池中的安培(阳极)检测器,该微流控池由微立体光刻制成,专门用于流动注射分析 (FIA)。流动池具有通道(管道)几何形状,高度为 25μm,显示出具有层流泊肃叶流动的良好水力定义。在溶液连续流过电极但电极仅短时间暴露于分析物的布置中,SWNT 电极不会被污染并且可以重复使用数月。本文报道的多巴胺 (DA) 的 LOD 明显低于以前使用 FIA-电化学检测的报告。此外,SWNT 电极可以原样使用,即它们不需要化学修饰或清洁。SWNT 电极的极低背景信号,由于稀疏的表面覆盖和 SWNTs 的低固有电容,意味着无需进行信号处理即可测量痕量水平下 DA 氧化的低电流。在人工脑液中也可以检测到 DA,在 50μL(约 2.5 fmol)中检测限 (LOD) 约为 50 pM。