Chemical and Biomolecular Engineering Department, UCLA, Los Angeles, CA, USA.
Biosens Bioelectron. 2013 Apr 15;42:256-60. doi: 10.1016/j.bios.2012.10.061. Epub 2012 Nov 6.
An implantable micromachined multi-electrode array (MEA) microprobe modified for utilization as a complete electrochemical biosensor for rapid glutamate detection is described. A post-fabrication method for electrochemical deposition of an iridium oxide (IrOx) film onto a designated microelectrode enabled incorporation of an IrOx reference electrode (RE) on the microprobe. The on-probe IrOx RE provides an alternative to the commonly utilized Ag/AgCl wire RE, which has been shown to be unstable and to cause an inflammatory response in living tissue. The IrOx film electrodeposited onto a platinum site was tested as part of a complete chemical sensing system that included a platinum counter electrode and enzymatic glutamate sensing electrodes all on a single silicon-based MEA platform. The thin film IrOx was mechanically robust enough to endure conditions of repeated heating and wetting during the MEA fabrication process. The pH dependence of the IrOx open circuit potential (OCP) was measured at -77±0.4 mV/pH and remained stable over a two-week period. The on-probe IrOx RE was tested in a two- and three-electrode system with glutamate biosensors. The biosensors were shown to detect a physiologically relevant range of glutamate concentrations and to reject the interferents, dopamine and ascorbic acid. By incorporating all of the electrodes onto a single device, baseline noise was reduced by an average of ∼61% in vitro and ∼71% in vivo with reduced tissue damage, since only a single probe needed to be implanted.
一种可植入的微机电系统(MEMS)多电极阵列(MEA)微探针经过修改,可作为完整的电化学生物传感器用于快速谷氨酸检测。描述了一种在制造后将氧化铱(IrOx)薄膜电化学沉积到指定微电极上的方法,从而可以在微探针上集成 IrOx 参比电极(RE)。与常用的 Ag/AgCl 丝 RE 相比,IrOx 参比电极具有稳定性和不会在活体组织中引起炎症反应的优势。沉积在铂位上的 IrOx 薄膜作为完整化学传感系统的一部分进行了测试,该系统包括铂对电极和酶促谷氨酸传感电极,均位于单个硅基 MEA 平台上。薄膜 IrOx 机械强度足够高,可以承受 MEA 制造过程中反复加热和润湿的条件。测量了 IrOx 开路电位(OCP)随 pH 值的变化,在 -77±0.4 mV/pH 下,在两周的时间内保持稳定。在具有谷氨酸生物传感器的两电极和三电极系统中测试了探针上的 IrOx RE。生物传感器被证明可以检测到生理相关的谷氨酸浓度范围,并拒绝干扰物多巴胺和抗坏血酸。通过将所有电极集成到单个设备上,与单独的探针植入相比,体内和体外的基线噪声分别平均降低了约 61%和 71%,同时减少了组织损伤。