Abstract

The purpose of this study was to investigate the accuracy of a new count-proportional method for the measurement of left ventricular volume when applied to gated equilibrium blood-pool imaging. An equation is developed that relates total chamber volume, Vt, to the area of a pixel (M) and the ratio (R) of total counts within the chamber to the counts within the hottest pixel in the chamber such that Vt = 1.38 M3R3/2. The value of M is a constant for the particular scintillation camera-collimator system and R is obtained from observed count rates. All calculated volumes were compared to volumes measured using biplane contrast ventriculography. In 25 patients, the method for ventricular volumes gave an r of 0.95 and an s.e.e. of 23 ml [Volume (nuclear) = 0.94 Volume (cath) + 1.3]. Endsystolic volume was best calculated from end-diastolic volume and ejection fraction. Manual regions of interest were more accurate than automated regions of interest. This method appears to be as accurate as more complex approaches and has the advantage of not requiring attenuation correction or blood sampling.

摘要

本研究的目的是探讨一种新的计数比例法在门控平衡血池成像中测量左心室容积时的准确性。推导了一个方程,该方程将总腔室容积Vt与像素面积(M)以及腔室内总计数与腔室内最热像素内计数的比值(R)相关联,即Vt = 1.38M³R³/²。M值对于特定的闪烁相机 - 准直器系统是一个常数,R可从观察到的计数率获得。将所有计算出的容积与使用双平面造影心室造影测量的容积进行比较。在25名患者中,心室容积测量方法的r值为0.95,标准误差为23 ml [核容积 = 0.94 × 造影容积 + 1.3]。收缩末期容积最好根据舒张末期容积和射血分数来计算。手动感兴趣区比自动感兴趣区更准确。该方法似乎与更复杂的方法一样准确,并且具有不需要衰减校正或血样采集的优点。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验