Abstract

PURPOSE

Accurate longitudinal relaxation time (T₁) of arterial blood is important in evaluating blood flow in tissue by arterial spin labeling magnetic resonance (MR) imaging. Few studies have reported the T₁ of human arterial blood in vivo, especially using 3-tesla MR imaging. T₁ values of human venous blood in vivo have been reported, but they differ from those measured in vitro. We aimed to evaluate the accurate T₁ of human arterial blood in vivo.

METHODS

We measured T₁ values of blood in 10 healthy volunteers in vivo using an inversion-recovery fast gradient-echo sequence and 3-tesla MR imaging unit. We also measured hematocrit (Hct) values of venous blood samples. After nonselective application of the inversion pulse using a body coil, we obtained MR imaging signals of arterial blood in the abdominal aorta. Similarly, we measured the signals of venous blood in the internal jugular vein. Inversion times varied between 200 and 5000 ms for imaging of the abdominal aorta and 200 and 2500 ms for imaging of the jugular vein. We also acquired signals without the inversion pulse. We estimated T₁ values from the data by nonlinear least squares fitting of a 3-parameter model.

RESULTS

The T value (mean±standard deviation) of arterial blood was 1779±80 ms and of venous blood, 1694±77 ms. The average Hct value was 0.47. The R₁ (=1/T₁) of arterial blood was related to the Hct value as: R₁=(0.59±0.16)Hct+(0.29±0.07) (mean±standard error) s⁻¹. For the venous blood, R₁=(0.70±0.11)Hct+(0.27±0.05) s⁻¹.

CONCLUSION

We observed a T₁ of human arterial blood in vivo of 1779±80 ms at a mean hematocrit value of 0.47 as determined by 3T MR imaging; an even longer T₁ value is expected with a hematocrit value less than 0.47.

摘要

目的

准确的动脉血液纵向弛豫时间(T₁)对于动脉自旋标记磁共振成像(MR)评估组织中的血流非常重要。很少有研究报道过人体动脉血液的 T₁ 值,特别是在使用 3 特斯拉磁共振成像时。虽然已经报道了人体静脉血液的 T₁ 值,但它们与体外测量的值不同。我们旨在评估人体动脉血液的准确 T₁ 值。

方法

我们使用反转恢复快速梯度回波序列和 3 特斯拉磁共振成像仪对 10 名健康志愿者体内的血液 T₁ 值进行了测量。我们还测量了静脉血样的血细胞比容(Hct)值。使用体线圈进行非选择性反转脉冲应用后,我们获得了腹部主动脉中动脉血液的 MR 成像信号。同样,我们测量了颈内静脉中静脉血的信号。用于腹部主动脉成像的反转时间在 200 到 5000 毫秒之间,用于颈内静脉成像的反转时间在 200 到 2500 毫秒之间。我们还获取了没有反转脉冲的信号。我们通过非线性最小二乘法拟合三参数模型来从数据中估计 T₁ 值。

结果

动脉血液的 T 值(平均值±标准偏差)为 1779±80 ms,静脉血液的 T 值为 1694±77 ms。平均 Hct 值为 0.47。动脉血液的 R₁(=1/T₁)与 Hct 值的关系为:R₁=(0.59±0.16)Hct+(0.29±0.07)(平均值±标准误差)s⁻¹。对于静脉血液,R₁=(0.70±0.11)Hct+(0.27±0.05)s⁻¹。

结论

我们在 3T MR 成像下观察到人体动脉血液的 T₁ 值为 1779±80 ms,平均血细胞比容值为 0.47;当血细胞比容值小于 0.47 时,T₁ 值预计会更长。