Comparing the Tonojet disposable tonometer with the traditional Goldmann tonometer in glaucomatous and non-glaucomatous eyes.

作者信息

Farrell S M J, Dooley I, O'Connell E, Bashir S, Foley-Nolan A, Kearns F, Logan P, Fulcher T

机构信息

Department of Ophthalmology, Beaumont Hospital, Beaumont Road, Dublin 9, Ireland.

出版信息

Int Ophthalmol. 2013 Aug;33(4):367-74. doi: 10.1007/s10792-012-9711-5. Epub 2013 Jan 23.

Abstract

Applanation tonometry is widely performed as the most accurate method of measuring intraocular pressure in the clinical setting. Disposable tonometer heads reduce the risk of transmitting infection during applanation and eliminate tonometer head cleaning, which is time-consuming. We performed a study to investigate whether the Tonojet disposable tonometer (Luneau, France) gave comparable intraocular pressure measurements to the standard Goldmann tonometer produced by Haag Streit (Switzerland). The study was performed in a busy general secondary referral ophthalmology clinic in Beaumont Hospital over a 15-day period. We measured intraocular pressure with the Goldmann and Tonojet tonometers in 293 eyes. Patients were randomised as to which tonometer was used first. A diagnosis of glaucoma or ocular hypertension was noted. Central corneal thickness (CCT) was measured. The measurements taken with the Tonojet were on average 2.2 mmHg lower than corresponding measurements taken with the Goldmann applanation tonometer head. The bias was present over the entire range of pressures. The 95 % limits of agreement between the two tonometers ranged from -5.9 to +1.4 mmHg. In the cohort of eyes with ocular hypertension/glaucoma the difference was significantly higher (2.8 vs. 2.1 mmHg). The bias was independent of CCT. We found that the Tonojet disposable tonometer tends to underestimate intraocular pressure compared with the Goldmann. The bias is non-systematic, present over the entire range of pressures and independent of corneal thickness. We conclude that the Tonojet is not sufficiently accurate to constitute a suitable alternative to the Goldmann.

摘要

压平眼压测量法作为临床环境中测量眼压最准确的方法被广泛应用。一次性眼压计探头降低了压平过程中传播感染的风险,并且无需进行耗时的眼压计探头清洁。我们进行了一项研究,以调查Tonojet一次性眼压计(法国Luneau公司)与Haag Streit(瑞士)生产的标准戈德曼眼压计测量的眼压是否具有可比性。该研究在博蒙特医院繁忙的普通二级转诊眼科诊所进行,为期15天。我们使用戈德曼眼压计和Tonojet眼压计测量了293只眼睛的眼压。患者被随机分配先使用哪种眼压计。记录青光眼或高眼压症的诊断情况。测量中央角膜厚度(CCT)。使用Tonojet眼压计测得的眼压平均比使用戈德曼压平眼压计探头测得的相应眼压低2.2 mmHg。在整个压力范围内均存在偏差。两种眼压计之间95%的一致性界限为-5.9至+1.4 mmHg。在高眼压症/青光眼患者组中,差异明显更高(2.8对2.1 mmHg)。该偏差与CCT无关。我们发现,与戈德曼眼压计相比,Tonojet一次性眼压计往往会低估眼压。该偏差是非系统性的,在整个压力范围内均存在,且与角膜厚度无关。我们得出结论,Tonojet眼压计的准确性不足以成为戈德曼眼压计的合适替代品。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验