新型直接血糖分析仪中使用全血作为样本材料的问题。

Problems with the use of whole blood as a sample material in novel direct glucose analysers.

作者信息

Savolainen K, Viitala A, Puhakainen E, Väisänen M

机构信息

Department of Clinical Chemistry, Kuopio University Central Hospital, Finland.

出版信息

Scand J Clin Lab Invest. 1990 Apr;50(2):221-3. doi: 10.1080/00365519009089158.

Abstract

We report a marked shift in whole-blood glucose measurements when a widely available glucose analyser is used. We compared a conventional glucose dehydrogenase (GDHG) method after protein precipitation and a direct polarographic method in conjunction with immobilized glucose oxidase. The polarographic method gave mean whole blood glucose levels from 79% to 95% of those measured by the GDHG method. No such difference was found in either plasma or serum samples or in water-based control samples. The stability of glucose in fluoride oxalate tubes appears to be poorer than is usually assumed. We observed a 5-10% decrease in whole-blood glucose concentrations during the first hour after sampling under routine conditions.

摘要

我们报告了在使用一种广泛应用的葡萄糖分析仪时全血葡萄糖测量值出现的显著变化。我们比较了蛋白质沉淀后传统的葡萄糖脱氢酶(GDHG)法和结合固定化葡萄糖氧化酶的直接极谱法。极谱法测得的全血葡萄糖平均水平为GDHG法测量值的79%至95%。在血浆或血清样本以及水性对照样本中均未发现此类差异。氟化物草酸盐试管中葡萄糖的稳定性似乎比通常认为的要差。我们观察到在常规条件下采样后的第一个小时内全血葡萄糖浓度下降了5%至10%。