医用雾化器产生的液滴。一些影响其大小和溶质浓度的因素。

Droplets produced by medical nebulizers. Some factors affecting their size and solute concentration.

作者信息

Phipps P R, Gonda I

机构信息

Department of Pharmacy, Sydney University, Australia.

出版信息

Chest. 1990 Jun;97(6):1327-32. doi: 10.1378/chest.97.6.1327.

Abstract

The effect of nebulizer solution temperature and dilution air humidity on the size and solute concentration of aqueous aerosol droplets were studied. Four combinations of jet-nebulizers with air compressors or oxygen sources and one ultrasonic nebulizer were tested. The temperature to which the nebulizer solution of each system fell during generation was measured. The nebulizers were then kept at set temperatures, generated aerosols collected and either droplet size or solute concentration measured. The droplet solute concentration was found to increase. The droplet size decreased along with the droplet solute concentration increase. The ultrasonic nebulizer also was tested: its high output made the concentration of the solution in the droplets much more stable. However, the proportion of droplets depositing in the tubing and valves changed markedly with aerosol flow rate. The potential for large changes in droplet solute concentration, droplet size and output during nebulization should be considered in therapeutic and diagnostic applications of nebulized aerosols.

摘要

研究了雾化器溶液温度和稀释空气湿度对水性气溶胶液滴大小和溶质浓度的影响。测试了四种将喷射雾化器与空气压缩机或氧气源组合的情况以及一种超声雾化器。测量了每个系统在产生过程中雾化器溶液下降到的温度。然后将雾化器保持在设定温度,收集产生的气溶胶并测量液滴大小或溶质浓度。发现液滴溶质浓度增加。液滴大小随着液滴溶质浓度的增加而减小。还对超声雾化器进行了测试:其高输出使液滴中溶液的浓度更加稳定。然而,沉积在管道和阀门中的液滴比例随气溶胶流速有明显变化。在雾化气溶胶的治疗和诊断应用中,应考虑雾化过程中液滴溶质浓度、液滴大小和输出发生大幅变化的可能性。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验