一项基于社区的Vita-Stat自动血压记录仪评估。

A community-based evaluation of the Vita-Stat automatic blood pressure recorder.

作者信息

Salaita K, Whelton P K, Seidler A J

机构信息

Department of Epidemiology, Johns Hopkins University School of Public Health and Hygiene, Baltimore, Maryland.

出版信息

Am J Hypertens. 1990 May;3(5 Pt 1):366-72. doi: 10.1093/ajh/3.5.366.

Abstract

A random sample of ten community-based Vita-Stat automatic blood pressure recorders (ABPR) was evaluated for accuracy and repeatability. Each of 518 adult volunteers had two randomly ordered systolic and diastolic Vita-Stat blood pressure measurements compared with two corresponding measurements obtained by a trained observer using a Random-Zero sphygmomanometer. Eight of the ten Vita-Stat recorders underestimated systolic blood pressure. In contrast, diastolic blood pressure was overestimated by nine of the ten machines. While the overall differences in average blood pressure were small (mean +/- SEM = -2.4 +/- 0.6 mm Hg for systolic blood pressure and +2.3 +/- 0.4 mm Hg for diastolic blood pressure), the magnitude of the average discrepancy varied considerably by machine (+4.7 to -13.8 mm Hg for systolic and +5.0 to -2.0 for diastolic blood pressure). At every level of systolic and diastolic blood pressure the Vita-Stat ABPR provided a less accurate method of classifying blood pressure among individuals than the human observer. Additional analyses exploring the ability of the Vita-Stat machine to measure an individual's blood pressure within 2, 4, or 6 mm Hg of the corresponding Random-Zero value again suggested that the Vita-Stat ABPR was less accurate than the human observer. Duplicate Vita-Stat readings were less repeatable than corresponding Random-Zero measurements. Based on the findings of this study, the Vita-Stat ABPR appears to be an unsatisfactory tool for community-based self-measurement of blood pressure.

摘要

对随机抽取的十台社区用Vita-Stat自动血压记录仪(ABPR)进行了准确性和重复性评估。518名成年志愿者每人都进行了两次随机排序的Vita-Stat收缩压和舒张压测量,并与经过培训的观察者使用随机零点血压计获得的两次相应测量结果进行比较。十台Vita-Stat记录仪中有八台低估了收缩压。相比之下,十台机器中有九台高估了舒张压。虽然平均血压的总体差异较小(收缩压的平均值±标准误=-2.4±0.6毫米汞柱,舒张压的平均值±标准误=+2.3±0.4毫米汞柱),但平均差异的幅度因机器而异(收缩压为+4.7至-13.8毫米汞柱,舒张压为+5.0至-2.0毫米汞柱)。在收缩压和舒张压的每个水平上,与人类观察者相比,Vita-Stat ABPR对个体血压分类的方法准确性较低。进一步分析Vita-Stat机器测量个体血压时与相应随机零点值相差2、4或6毫米汞柱以内的能力,结果再次表明Vita-Stat ABPR的准确性低于人类观察者。Vita-Stat的重复读数比相应的随机零点测量结果重复性更差。基于本研究的结果,Vita-Stat ABPR似乎是一种不适合社区自我测量血压的工具。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验