Matsumoto Tetsuro, Harano Hideki, Masuda Akihiko, Nishiyama Jun, Matsue Hideaki, Uritani Akira, Nunomiya Tomoya
National Institute of Advanced Industrial Science and Technology, Tsukuba 305-8568, Japan.
Radiat Prot Dosimetry. 2013 Aug;155(4):505-11. doi: 10.1093/rpd/nct029. Epub 2013 Mar 17.
A new thermal neutron calibration method to experimentally determine the energy response function of a neutron detector using a pulse parallel beam and the time-of-flight (TOF) technique is developed. The calibration method was experimentally demonstrated for a (3)He proportional counter and an electric personal dosemeter using a pulsed thermal neutron beam from the research reactor JRR-3M. The responses of the detectors were successfully obtained as a function of neutron energy. However, detailed information on the detector structure is required to obtain the spatial response distribution for the detector. The authors further propose an improved calibration method obtaining the spatial response distribution using a pulsed narrow beam, the TOF technique and a beam scanning technique.
开发了一种新的热中子校准方法,用于使用脉冲平行束和飞行时间(TOF)技术通过实验确定中子探测器的能量响应函数。利用来自研究堆JRR-3M的脉冲热中子束,对³He正比计数器和个人电子剂量计进行了该校准方法的实验验证。成功获得了探测器响应随中子能量的变化关系。然而,要获得探测器的空间响应分布,需要有关探测器结构的详细信息。作者进一步提出了一种改进的校准方法,使用脉冲窄束、TOF技术和束扫描技术来获得空间响应分布。