The orientation of the acetabular component in metal-on-metal hip resurfacing arthroplasty affects wear rate and hence failure. This study aimed to establish if interpretation of pelvic radiographs with TraumaCad software can provide a reliable alternative to CT in measuring the acetabular inclination and version.
TraumaCad was used to measure the acetabular orientation on AP pelvis radiographs of 14 painful hip resurfacings. Four orthopaedic surgeons performed each measurement twice. These were compared with measurements taken from CT reformats. The correlation between TraumaCad and CT was calculated, as was the intra- and inter-observer reliability of TraumaCad.
There is strong correlation between the two techniques for the measurement of inclination and version (p <0.001). Intra- and inter-observer reliability of TraumaCad measurements are good (p <0.001). Mean absolute error for measurement of inclination was 2.1°. TraumaCad underestimated version compared to CT in 93% of cases, by 12.6 degrees on average.
When assessing acetabular orientation in hip resurfacing, the orthopaedic surgeon may use TraumaCad in the knowledge that it correlates well with CT and has good intra- and inter-observer reliability but underestimates version by 12° on average.
金属对金属髋关节表面置换术中髋臼部件的方向会影响磨损率,从而影响假体的失败。本研究旨在确定 TraumaCad 软件是否可以用于骨盆平片来替代 CT 来测量髋臼的倾斜度和前倾角。
使用 TraumaCad 软件测量 14 例髋关节置换术后出现疼痛的髋关节正位骨盆平片上髋臼的位置。4 名矫形外科医生分别进行了两次测量。这些测量值与 CT 重建值进行了比较。计算了 TraumaCad 与 CT 之间的相关性,以及 TraumaCad 的内部和外部观察者可靠性。
两种技术在测量倾斜度和前倾角方面具有很强的相关性(p <0.001)。TraumaCad 测量的内部和外部观察者可靠性均较好(p <0.001)。倾斜度测量的平均绝对误差为 2.1°。在 93%的情况下,TraumaCad 与 CT 相比低估了髋臼前倾角,平均低估了 12.6 度。
在评估髋关节表面置换的髋臼位置时,矫形外科医生可以使用 TraumaCad,因为它与 CT 具有很好的相关性,且内部和外部观察者可靠性均较好,但平均低估了髋臼前倾角 12 度。