Alario Anthony F, Pirie Christopher G
Department of Clinical Sciences, Tufts Cummings School of Veterinary Medicine, North Grafton, MA, 01536, USA.
Vet Ophthalmol. 2014 May;17(3):207-11. doi: 10.1111/vop.12074. Epub 2013 Jun 14.
To compare central corneal thickness (CCT) measurements obtained from normal canines using a spectral-domain optical coherence tomography device (SD-OCT) and an ultrasonic pachymeter.
Fifteen clinically normal dogs.
Dogs were briefly anesthetized with propofol. Central corneal thickness measurements were obtained from both eyes of each animal four times by one operator using a SD-OCT system. The CCT was automatically calculated using the pachymetry software generated from eight radial scans, 6 mm in length. Ultrasonic pachymetry was then performed in both eyes of each animal by one operator five times in succession. The CCT was recorded for each modality and compared statistically for agreement and bias.
Fifteen dogs (seven castrated male, eight spayed female) with a mean age of 2.3 ± 1.3 years were examined. The mean CCT for all eyes examined via SD-OCT and ultrasonic pachymetry (velocity set at 1636 m/s) was 587.72 ± 32.44 μm and 598.54 ± 32.28 μm, respectively. Ultrasonic pachymetry consistently overestimated CCT as compared to OCT by a mean value of 10.82 μm (P < 0.0001). No significant difference in CCT was found based on the age or sex of the animal. There was no significant difference in CCT measurements between replicates performed by the same operator for each modality. The coefficient of variation within and between imaging modalities was low (range 0.63-2.5%). The intraclass correlation coefficient comparing methods was 0.92.
Correlation between modalities was excellent; however, ultrasonic pachymetry set at a standard velocity of 1636 m/s overestimates CCT as compared to OCT.
使用光谱域光学相干断层扫描设备(SD-OCT)和超声测厚仪比较从正常犬获得的中央角膜厚度(CCT)测量值。
15只临床正常的犬。
用丙泊酚对犬进行短暂麻醉。由一名操作人员使用SD-OCT系统对每只动物的双眼进行4次中央角膜厚度测量。使用由8次长度为6mm的径向扫描生成的测厚软件自动计算CCT。然后由一名操作人员对每只动物的双眼连续进行5次超声测厚。记录每种方法的CCT,并进行统计学比较以评估一致性和偏差。
检查了15只犬(7只去势雄性,8只绝育雌性),平均年龄为2.3±1.3岁。通过SD-OCT和超声测厚仪(速度设定为1636m/s)检查的所有眼睛的平均CCT分别为587.72±32.44μm和598.54±32.28μm。与OCT相比,超声测厚仪始终高估CCT,平均值为10.82μm(P < 0.0001)。未发现动物的年龄或性别对CCT有显著影响。同一操作人员对每种方法进行的重复测量之间的CCT测量值无显著差异。成像方法内部和之间的变异系数较低(范围为0.63 - 2.5%)。比较方法的组内相关系数为0.92。
两种方法之间的相关性极佳;然而,与OCT相比,速度设定为1636m/s的标准超声测厚仪高估了CCT。