Comparison of ultrasonic pachymetry and Fourier-domain optical coherence tomography for measurement of corneal thickness in dogs with and without corneal disease.

作者信息

Hoehn A L, Thomasy S M, Kass P H, Horikawa T, Samuel M, Shull O R, Stewart K A, Murphy C J

机构信息

Department of Surgical and Radiological Sciences, School of Veterinary Medicine, University of California, Davis, One Shields Avenue, Davis, CA 95616, USA.

Department of Surgical and Radiological Sciences, School of Veterinary Medicine, University of California, Davis, One Shields Avenue, Davis, CA 95616, USA; Department of Ophthalmology & Vision Science, School of Medicine, University of California, Davis, 4610 X St., Sacramento, CA 95817, USA.

出版信息

Vet J. 2018 Dec;242:59-66. doi: 10.1016/j.tvjl.2018.10.008. Epub 2018 Oct 30.

Abstract

Several ultrasonic and Fourier-domain optical coherence tomography (FD-OCT) pachymeters are used to measure corneal thickness in canine patients and research subjects. This study assessed the reliability of and consistency between two ultrasonic pachymetry (USP) devices, Pachette 3 and Accupach VI, as well as automated and manual measurements obtained using FD-OCT in dogs with and without corneal disease. Corneal thickness measurements were compiled from 108 dogs and analyzed using mixed effects linear regression, with Bonferonni adjustments for post-hoc comparisons, to determine the effects of age, weight and disease state. Data are presented as predicted mean±standard error. Canine corneal disease can result in marked increases in thickness that frequently exceed the upper limits of measurement of some pachymetry devices developed for human use. In this study, the corneas of dogs with endothelial disease or injury frequently exceeded the upper limits of quantitation of 999 and 800μm for the Accupach VI and automated FD-OCT pachymeters, respectively. Using values <800μm, the Pachette 3 generated significantly greater values for central corneal thickness (CCT) than the Accupach VI, manual FD-OCT and automated FD-OCT at 625±7.0, 615±7.2, 613±7.2, and 606±7.4μm respectively (P<0.001). Of the two devices where measurements >1000μm were obtained, manual FD-OCT demonstrated less variability than the Pachette 3. Corneal thickness increased linearly with age and weight with an increase of 6.9±1.8μm/year and 1.6±0.8μm/kg body weight (P<0.005 and P=0.038, respectively).

摘要

几种超声和傅里叶域光学相干断层扫描(FD-OCT)测厚仪用于测量犬科患者和研究对象的角膜厚度。本研究评估了两种超声测厚仪(USP)——Pachette 3和Accupach VI之间的可靠性和一致性,以及在患有和未患有角膜疾病的犬中使用FD-OCT进行自动和手动测量的情况。对108只犬的角膜厚度测量数据进行汇总,并使用混合效应线性回归分析,采用Bonferonni校正进行事后比较,以确定年龄、体重和疾病状态的影响。数据以预测平均值±标准误差表示。犬角膜疾病可导致厚度显著增加,经常超过一些为人类开发的测厚仪的测量上限。在本研究中,患有内皮疾病或损伤的犬的角膜,分别经常超过Accupach VI和自动FD-OCT测厚仪999μm和800μm的定量上限。使用小于800μm的值时,Pachette 3测量的中央角膜厚度(CCT)值显著高于Accupach VI、手动FD-OCT和自动FD-OCT,分别为625±7.0、615±7.2、613±7.2和606±7.4μm(P<0.001)。在获得测量值>1000μm的两种设备中,手动FD-OCT的变异性小于Pachette 3。角膜厚度随年龄和体重呈线性增加,每年增加6.9±1.8μm,每千克体重增加1.6±0.8μm(分别为P<0.005和P=0.038)。