Intra and inter-user reliability of central corneal thickness measurements obtained in healthy feline eyes using a portable spectral-domain optical coherence tomography device.

作者信息

Alario Anthony F, Pirie Christopher G

机构信息

Department of Clinical Sciences, Tufts Cummings School of Veterinary Medicine, North Grafton, MA, 01536, USA.

出版信息

Vet Ophthalmol. 2013 Nov;16(6):446-50. doi: 10.1111/vop.12029. Epub 2013 Jan 29.

Abstract

OBJECTIVE

To determine the intra and inter-user reliability of central corneal thickness (CCT) measurements obtained using a portable spectral-domain optical coherence tomography (SD-OCT) device in healthy feline eyes.

ANIMALS

Twenty clinically normal cats.

PROCEDURES

Gentle manual restraint was utilized for proper animal positioning. Central corneal thickness measurements were obtained from both eyes of each animal three times by two operators in succession. A 6 mm, corneal pachymetry protocol was performed using the Optovue® iVue SD-OCT system. The CCT was automatically calculated using the pachymetry software generated from eight radial scans, 6 mm in length. The CCT was recorded for each operator and compared statistically for both intra and inter-user reliability.

RESULTS

Twenty cats (11 CM, 9 SF) with a mean age of 4.7 years were examined. The mean CCT for all eyes examined was 629.08 ± 47.05 μm. No significant difference in CCT was found based on the eye examined (OD vs. OS) or age of the animal. However, the CCT of spayed females was significantly higher (42.64 ± 18.0 μm) as compared to castrated males (P = 0.03). There was no significant difference in CCT measurements between replicates performed by the same operator or between operators. The mean difference in CCT between operators was 0.38 μm. The coefficient of variation for each user and between users was very low (range 0.68-1.5%). The intra-class correlation coefficient comparing operators was 0.975.

CONCLUSIONS

The Optovue® iVue SD-OCT device is capable of obtaining CCT measurements with excellent intra and interoperator reliability in healthy feline eyes.

摘要

目的

确定使用便携式光谱域光学相干断层扫描(SD - OCT)设备在健康猫眼中获得的中央角膜厚度(CCT)测量值的使用者内和使用者间可靠性。

动物

20只临床正常的猫。

方法

采用轻柔的手动约束以正确定位动物。两名操作人员依次对每只动物的双眼进行三次中央角膜厚度测量。使用Optovue® iVue SD - OCT系统执行6毫米角膜测厚方案。使用由8条长度为6毫米的径向扫描生成的测厚软件自动计算CCT。记录每位操作人员的CCT,并对使用者内和使用者间可靠性进行统计学比较。

结果

检查了20只猫(11只混种猫,9只短毛猫),平均年龄为4.7岁。所有检查眼睛的平均CCT为629.08±47.05微米。根据检查的眼睛(右眼与左眼)或动物年龄,未发现CCT有显著差异。然而,绝育雌性猫的CCT显著高于去势雄性猫(42.64±18.0微米,P = 0.03)。同一操作人员进行的重复测量之间或操作人员之间的CCT测量无显著差异。操作人员之间CCT的平均差异为0.38微米。每位使用者以及使用者之间的变异系数非常低(范围0.68 - 1.5%)。比较操作人员的组内相关系数为0.975。

结论

Optovue® iVue SD - OCT设备能够在健康猫眼中获得具有出色使用者内和使用者间可靠性的CCT测量值。