Jingjing Zhang, Jie Jiang, Xiaolin Qiao
School of Information and Electrical Engineering, Harbin Institute of Technology at Weihai, 2 West Wenhua Road, Weihai 264209, People's Republic of China.
Rev Sci Instrum. 2013 Jun;84(6):065103. doi: 10.1063/1.4807746.
Paper spray (PS) ionization is a recently developed ion source that has been used to analyze samples in their native environments at ambient pressure without requiring sample preparation or pre-separation. The design of an ion mobility spectrometer (IMS) coupled with PS ionization can help expand IMS applications to on-site detection of complex liquid samples. We report a paper spray ionization ion mobility spectrometer prototype that consists of a PS source and an ion mobility spectrometer optimized using a numerical simulation. The performance of the design was evaluated by measuring 2, 6-di-tert-butylpyridine (2, 6-DtBP). The mobility spectra of the 2, 6-DtBP exhibited a single-product ion peak with reduced mobility calculated at 1.42 cm(2)∕(V s) and a linear response of 0.1-10 μg∕ml, with an estimated detection limit of 0.05 μg∕ml. The Relative Standard Deviation for 1 μg∕ml was 5.7% over 11 measurements. The highest resolving power (47) was measured for 2, 6-DtBP. Based on these preliminary results, the present PSIMS design is expected to become a tool of choice for the rapid analysis of complex liquid samples.
纸喷雾(PS)电离是一种最近开发的离子源,已用于在常压下分析其天然环境中的样品,无需样品制备或预分离。与PS电离相结合的离子迁移谱仪(IMS)的设计有助于将IMS应用扩展到复杂液体样品的现场检测。我们报告了一种纸喷雾电离离子迁移谱仪原型,它由一个PS源和一个通过数值模拟优化的离子迁移谱仪组成。通过测量2,6-二叔丁基吡啶(2,6-DtBP)来评估该设计的性能。2,6-DtBP的迁移谱显示出一个单产物离子峰,其折合迁移率计算为1.42 cm²∕(V·s),线性响应范围为0.1 - 10 μg∕ml,估计检测限为0.05 μg∕ml。在11次测量中,1 μg∕ml的相对标准偏差为5.7%。对2,6-DtBP测量到的最高分辨率为47。基于这些初步结果,目前的PSIMS设计有望成为快速分析复杂液体样品的首选工具。