Suppr超能文献

测量瑞士初级保健医生医疗实践中血压计的准确性。

Measuring accuracy of sphygmomanometers in the medical practices of Swiss primary care physicians.

机构信息

Department of Cardiology, Cantonal Hospital of Lucerne , Lucerne , Switzerland.

出版信息

Eur J Gen Pract. 2013 Dec;19(4):244-7. doi: 10.3109/13814788.2013.779664. Epub 2013 Jul 3.

Abstract

OBJECTIVE

Arterial hypertension has a high prevalence in most countries. Blood pressure measurements are performed frequently by primary care physicians. Recommendations from different societies emphasise the importance of measuring blood pressure with well maintained and calibrated instruments only. Since appropriate quality control measures are lacking the following survey was conducted in the medical practices of Swiss primary care physicians.

METHODS

This is a cross-sectional survey with Swiss primary care physicians. Nine hundred and seventy-five sphygmomanometers used in the daily practice of medicine were compared and calibrated against a certified calibrator. The magnitude of the measuring error before and after calibration was determined.

RESULTS

The proportion of the instruments that measured within the required tolerance of ± 3 mmHg over all measuring ranges was 81.4%. The average maintenance time was 5.6 years (± 3.8), and 97% (n = 353) of these instruments had not been maintained for two years (i.e. the recommended maintenance interval) or more. Two years after maintenance the number of devices with measurement errors of more than ± 3 mmHg increased significantly.

CONCLUSION

In Swiss primary care practices, the majority of upper arm and wrist sphygmomanometers measured blood pressure within a tolerance of ± 3 mmHg despite low adherence to the recommended maintenance interval. Two years after maintenance the number of sphygmomanometers with measurement errors increased significantly.

摘要

目的

在大多数国家,高血压的患病率都很高。初级保健医生经常进行血压测量。来自不同协会的建议强调了仅使用维护和校准良好的仪器测量血压的重要性。由于缺乏适当的质量控制措施,因此对瑞士初级保健医生的医疗实践进行了以下调查。

方法

这是一项针对瑞士初级保健医生的横断面调查。将日常医学实践中使用的 975 台血压计与经过认证的校准器进行了比较和校准。在进行和未进行校准之前,确定了测量误差的大小。

结果

在所测的所有测量范围内,测量误差在±3mmHg 以内的仪器比例为 81.4%。平均维护时间为 5.6 年(±3.8),其中 97%(n=353)的仪器未达到两年(即推荐的维护间隔)或更长时间进行维护。维护两年后,测量误差超过±3mmHg 的设备数量显著增加。

结论

在瑞士的初级保健实践中,尽管大多数上臂和手腕血压计都在±3mmHg 的容差范围内测量血压,但大多数设备都没有遵守推荐的维护间隔。维护两年后,测量误差的血压计数量显著增加。