IEEE Trans Biomed Circuits Syst. 2011 Apr;5(2):147-59. doi: 10.1109/TBCAS.2011.2138139.
This paper presents a multichannel stimulator ASIC for an implantable vestibular prosthesis. The system features versatile stimulation management which allows fine setting of the parameters for biphasic stimulation pulses. To address the problem of charge imbalance due to rounding errors, the digital processor can calculate and provide accurate charge correction. A technique to reduce the data rate to the stimulator is described. The stimulator ASIC was implemented in 0.6-μ m high-voltage CMOS technology occupying an area of 2.27 mm(2). The measured performance of the ASIC has been verified using vestibular electrodes in saline.
本文介绍了一种用于可植入前庭假体的多通道刺激器 ASIC。该系统具有多功能刺激管理功能,可精细设置双相刺激脉冲的参数。为了解决由于舍入误差导致的电荷失衡问题,数字处理器可以计算并提供精确的电荷校正。本文还描述了一种降低刺激器数据速率的技术。该刺激器 ASIC 采用 0.6μm 高压 CMOS 技术实现,占用面积为 2.27mm²。使用盐水中的前庭电极对 ASIC 的测量性能进行了验证。