Department of Psychology, Université de Montréal, Canada.
J Clin Sleep Med. 2013 Jul 15;9(7):701-6. doi: 10.5664/jcsm.2844.
The algorithms used to derive sleep variables from actigraphy were developed with adults. Because children change position during sleep more often than adults, algorithms may detect wakefulness when the child is actually sleeping (false negative). This study compares the validity of three algorithms for detecting sleep with actigraphy by comparing them to PSG in preschoolers. The putative influence of device location (wrist or ankle) is also examined.
Twelve children aged 2 to 5 years simultaneously wore an actigraph on an ankle and a wrist (Actiwatch-L, Mini-Mitter/Respironics) during a night of PSG recording at home. Three algorithms were tested: one recommended for adults and two designed to decrease false negative detection of sleep in children.
Actigraphy generally showed good sensitivity (> 95%; PSG sleep detection) but low specificity (± 50%; PSG wake detection). Intraclass correlations between PSG and actigraphy variables were strong (> 0.80) for sleep latency, sleep duration, and sleep efficiency, but weak for number of awakenings (< 0.40). The two algorithms designed for children enhanced the validity of actigraphy in preschoolers and increased the proportion of actigraphy-scored wake epochs scored that were also PSG-identified as wake. Sleep variables derived from the ankle and wrist were not statistically different.
Despite the weak detection of wakefulness, Acti-watch-L appears to be a useful instrument for assessing sleep in preschoolers when used with an adapted algorithm.
从动作记录仪中得出睡眠变量的算法是为成年人开发的。由于儿童在睡眠中更频繁地变换姿势,因此算法可能会错误地将儿童的睡眠状态识别为清醒(假阴性)。本研究通过与 PSG 比较,比较了三种动作记录仪检测学龄前儿童睡眠的算法的有效性。还检查了设备位置(手腕或脚踝)的潜在影响。
12 名 2 至 5 岁的儿童同时在家庭 PSG 记录期间在脚踝和手腕上佩戴动作记录仪(Actiwatch-L,Mini-Mitter/Respironics)。测试了三种算法:一种推荐用于成年人,另外两种旨在减少儿童睡眠状态的假阴性检测。
动作记录仪通常具有较高的敏感性(>95%;PSG 睡眠检测),但特异性较低(±50%;PSG 清醒检测)。PSG 和动作记录仪变量之间的组内相关系数很强(>0.80),用于睡眠潜伏期、睡眠时间和睡眠效率,但对于觉醒次数较弱(<0.40)。为儿童设计的两种算法提高了动作记录仪在学龄前儿童中的有效性,并增加了动作记录仪评分的清醒阶段比例,这些阶段也被 PSG 识别为清醒。脚踝和手腕得出的睡眠变量在统计学上没有差异。
尽管清醒状态的检测较弱,但当使用适应性算法时,Acti-watch-L 似乎是评估学龄前儿童睡眠的有用工具。