利用超声背散射系数评估高强度聚焦超声治疗鼠类乳腺肿瘤。

Assessment of high-intensity focused ultrasound treatment of rodent mammary tumors using ultrasound backscatter coefficients.

机构信息

Bioacoustics Research Laboratory, Department of Electrical and Computer Engineering, University of Illinois at Urbana-Champaign, 405 North Mathews, Urbana, Illinois 61081, USA.

出版信息

J Acoust Soc Am. 2013 Aug;134(2):1559-68. doi: 10.1121/1.4812877.

Abstract

Fischer 344 rats with subcutaneous mammary adenocarcinoma tumors were exposed to therapeutic ultrasound at one of three exposure levels (335, 360, and 502 W/cm(2) spatial-peak temporal-average intensity). Quantitative ultrasound estimates were generated from ultrasound radio frequency (RF) data from tumors before and after high-intensity focused ultrasound treatment. Treatment outcome was independently assessed by triphenyl tetrazolium chloride (TTC) staining, histological analysis by a pathologist, and thermocouple data. The average backscatter coefficient (BSC) and integrated backscatter coefficient (IBSC) were estimated before and after therapeutic ultrasound exposure for each tumor from RF data collected using clinical (Ultrasonix Sonix RP) and small-animal (Visualsonics Vevo 2100) array systems. Changes in the BSC with treatment were comparable to inter-sample variation of untreated tumors, but statistically significant differences in the change in the IBSCs were observed when comparing the exposures collectively (p < 0.10 for Sonix RP, p < 0.05 for Vevo 2100). Several exposure levels produced statistically significant differences in the change in IBSC when examined pair-wise, including two exposures having similar intensities (p < 0.05, Vevo 2100). A comparison of the IBSC results with temperature data, histology, and TTC staining revealed that the BSC was not always sensitive to thermal insult and that peak exposure pressure appeared to correlate with observed BSC increases.

摘要

患有皮下乳腺腺癌肿瘤的 Fischer 344 大鼠分别在三种超声辐照强度(335、360 和 502W/cm2 空间峰值-时间平均强度)下接受治疗性超声辐照。高强度聚焦超声治疗前后,从肿瘤的超声射频(RF)数据中生成了定量超声估计值。治疗结果通过三苯基四唑氯(TTC)染色、病理学家的组织学分析和热电偶数据独立评估。治疗性超声辐照前后,使用临床(Ultrasonix Sonix RP)和小动物(Visualsonics Vevo 2100)阵列系统收集的 RF 数据,从每个肿瘤估算平均背散射系数(BSC)和积分背散射系数(IBSC)。治疗后 BSC 的变化与未处理肿瘤的样本间变化相当,但当比较所有暴露时,IBSC 的变化存在统计学显著差异(Sonix RP 为 p<0.10,Vevo 2100 为 p<0.05)。当逐一比较时,几个辐照水平导致 IBSC 变化存在统计学显著差异,包括两个强度相似的辐照(Vevo 2100,p<0.05)。将 IBSC 结果与温度数据、组织学和 TTC 染色进行比较表明,BSC 并不总是对热损伤敏感,并且峰值辐照压力似乎与观察到的 BSC 增加相关。

相似文献

3
Quantitative Ultrasound for Monitoring High-Intensity Focused Ultrasound Treatment In Vivo.超声弹性成像技术监测高强度聚焦超声治疗的体内疗效。
IEEE Trans Ultrason Ferroelectr Freq Control. 2016 Sep;63(9):1234-42. doi: 10.1109/TUFFC.2016.2517644. Epub 2016 Jan 14.

本文引用的文献

1
Ex vivo study of quantitative ultrasound parameters in fatty rabbit livers.兔脂肪肝的超声定量参数的离体研究。
Ultrasound Med Biol. 2012 Dec;38(12):2238-48. doi: 10.1016/j.ultrasmedbio.2012.08.010. Epub 2012 Oct 11.
2
Ultrasonic assessment of thermal therapy in rat liver.超声评估大鼠肝脏热疗。
Ultrasound Med Biol. 2012 Dec;38(12):2130-7. doi: 10.1016/j.ultrasmedbio.2012.07.024. Epub 2012 Oct 11.
4
Three-dimensional high-frequency characterization of cancerous lymph nodes.三维高频特征分析在癌性淋巴结中的应用。
Ultrasound Med Biol. 2010 Mar;36(3):361-75. doi: 10.1016/j.ultrasmedbio.2009.10.007. Epub 2010 Feb 4.
8
Two-dimensional temperature estimation using diagnostic ultrasound.使用诊断超声进行二维温度估计。
IEEE Trans Ultrason Ferroelectr Freq Control. 1998;45(4):1088-99. doi: 10.1109/58.710592.
9
Treatment monitoring and thermometry for therapeutic focused ultrasound.
Int J Hyperthermia. 2007 Mar;23(2):121-39. doi: 10.1080/02656730701207842.