A study of the effects of sodium halides on the performance of air-jet and vibrating-mesh nebulizers.

机构信息

Faculty of Pharmacy, Albaath University, Homs, Syria.

出版信息

Int J Pharm. 2013 Nov 18;456(2):520-7. doi: 10.1016/j.ijpharm.2013.08.023. Epub 2013 Aug 21.

Abstract

The influence of sodium halide electrolytes on aerosols generated from the Aeroneb Pro vibrating mesh nebulizer and the Sidestream air-jet nebulizer has been evaluated. Fluids with a range of concentrations of Na halides (i.e. NaF, NaCl, NaBr and NaI) were used as nebulizer solutions and their effect on aerosol properties such as total aerosol output, fine particle fraction (FPF), volume median diameter (VMD) and predicted regional airway deposition were investigated. For both nebulizers, the inclusion of electrolyte significantly enhanced the aerosol properties compared with HPLC grade (deionized) water. Aerosol output, FPF and aerosol fraction less than 2.15 μm were directly proportional to electrolyte concentration. Furthermore, the proportion of aerosols that are likely to deposit in the oropharyngeal region, and the VMD of the droplets were inversely related to the electrolyte concentration for both nebulizers. In general, the inclusion of electrolytes had a greater impact on the aerosol properties of the vibrating-mesh nebulizer. In the Aeroneb Pro, NaI 2.0% (w/v) was the optimum solution as it generated the highest aerosol output, FPF and output fraction below 2.15 μm with the lowest VMD and minimal predicted oropharyngeal deposition. This was attributed to the polarizing ability of iodide ions present in the largest quantity at the air-water interface. This study has shown that the Aeroneb Pro vibrating-mesh device demonstrated greatly enhanced aerosol properties when halides were included in the nebulizer solutions.

摘要

评估了卤化钠电解质对 Aeroneb Pro 振动网式雾化器和 Sidestream 射流式雾化器产生的气溶胶的影响。使用了一系列浓度的 Na 卤化物(即 NaF、NaCl、NaBr 和 NaI)作为雾化器溶液,并研究了它们对气溶胶特性(如总气溶胶输出、细颗粒分数(FPF)、体积中值直径(VMD)和预测的区域性气道沉积)的影响。对于两种雾化器,与 HPLC 级(去离子)水相比,电解质的存在显著增强了气溶胶特性。气溶胶输出、FPF 和小于 2.15μm 的气溶胶分数与电解质浓度成正比。此外,对于两种雾化器,口咽区域沉积的气溶胶比例和液滴的 VMD 与电解质浓度成反比。一般来说,电解质的存在对振动网式雾化器的气溶胶特性影响更大。在 Aeroneb Pro 中,2.0%(w/v)的 NaI 是最佳溶液,因为它产生了最高的气溶胶输出、FPF 和小于 2.15μm 的输出分数,同时具有最低的 VMD 和最小的预测口咽沉积。这归因于界面处数量最多的碘离子的极化能力。本研究表明,当雾化器溶液中包含卤化物时,Aeroneb Pro 振动网式装置表现出大大增强的气溶胶特性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验