Joos Karen M, Shen Jin-Hui
Vanderbilt Eye Institute, Vanderbilt University, 2311 Pierce Avenue, Nashville, TN 37232, USA.
Biomed Opt Express. 2013 Jul 16;4(8):1342-50. doi: 10.1364/BOE.4.001342. eCollection 2013.
Optical coherence tomography (OCT) has a tremendous global impact upon the ability to diagnose, treat, and monitor eye diseases. A miniature 25-gauge forward-imaging OCT probe with a disposable tip was developed for real-time intraoperative ocular imaging of posterior pole and peripheral structures to improve vitreoretinal surgery. The scanning range was 2 mm when the probe tip was held 3-4 mm from the tissue surface. The axial resolution was 4-6 µm and the lateral resolution was 25-35 µm. The probe was used to image cellophane tape and multiple ocular structures.
光学相干断层扫描(OCT)对眼部疾病的诊断、治疗和监测能力产生了巨大的全球影响。为了改善玻璃体视网膜手术,研发了一种带有一次性探头尖端的微型25G前向成像OCT探头,用于后极部和周边结构的实时术中眼部成像。当探头尖端距组织表面3 - 4毫米时,扫描范围为2毫米。轴向分辨率为4 - 6微米,横向分辨率为25 - 35微米。该探头用于对玻璃纸胶带和多个眼部结构进行成像。