Sun Cuiru, Lee Kenneth K C, Vuong Barry, Cusimano Michael D, Brukson Alexander, Mauro Antonio, Munce Nigel, Courtney Brian K, Standish Beau A, Yang Victor X D
Biomed Opt Express. 2012 Jun 1;3(6):1404-12. doi: 10.1364/BOE.3.001404. Epub 2012 May 16.
A prototype intraoperative hand-held optical coherence tomography (OCT) imaging probe was developed to provide micron resolution cross-sectional images of subsurface tissue during open surgery. This new ergonomic probe was designed based on electrostatically driven optical fibers, and packaged into a catheter probe in the form factor of clinically accepted Bayonet shaped neurosurgical probes. Optical properties of the probe were measured to have a ~20 μm spot size, 5 mm working distance and 4 mm field of view. Feasibility of this probe for structural and Doppler shift imaging was tested on porcine femoral blood vessel imaging.