Suppr超能文献

通过二维剪切波弹性成像(2D-SWE)评估肝脏硬度需要进行多少次测量,以及应该使用哪个值:平均值还是中位数?

How many measurements are needed for liver stiffness assessment by 2D-Shear Wave Elastography (2D-SWE) and which value should be used: the mean or median?

作者信息

Sporea Ioan, Grădinaru-Taşcău Oana, Bota Simona, Popescu Alina, Şirli Roxana, Jurchiş Ana, Popescu Mădălina, Dănilă Mirela

机构信息

Department of Gastroenterology and Hepatology, "Victor Babes" University of Medicine and Pharmacy, Timisoara, Romania; Email:

出版信息

Med Ultrason. 2013 Dec;15(4):268-72. doi: 10.11152/mu.2013.2066.154.isp2.

Abstract

AIM

To analyze how many measurements are needed for non-invasive assessment of liver stiffness (LS) by means of 2D-Shear Wave Elastography (2D-SWE) and if the use of mean of median values of 2D-SWE measurements is needed.

METHODS

We evaluated 449 consecutive subjects (with or without chronic liver disease) by means of Transient Elastography (TE) and 2D-SWE. We compared the correlation of LS assessed by TE with 2D-SWE measurements when using either the median of 5 valid 2D-SWE measurements or the mean of 3 or 5 valid 2D-SWE measurements.

RESULTS

We obtained reliable LS measurements by TE in 330/449 subjects (73.5%). From these, in 281 subjects we obtained 5 valid 2D-SWE measurements. The correlation of LS assessed by TE with 2D-SWE values was similar when we used the median value of 5 valid 2D-SWE measurements, the mean value of 5 valid 2D-SWE measurements or the mean value of 3 valid 2D-SWE measurements: r =0.683, r=0.711 and r=0.691, respectively. There were no significant differences between the median value of 5 valid 2D-SWE measurements; the mean value of 5 valid 2D-SWE measurements; or the mean value of 3 valid 2D-SWE measurements: 7.6 kPa, 7.7 kPa and 7.6 kPa, respectively.

CONCLUSIONS

Our study showed that it is enough to perform 3 valid 2D-SWE measurements and to use the mean value of these measurements.

摘要

目的

分析通过二维剪切波弹性成像(2D-SWE)对肝脏硬度(LS)进行无创评估需要进行多少次测量,以及是否需要使用2D-SWE测量值的中位数均值。

方法

我们通过瞬时弹性成像(TE)和2D-SWE对449例连续受试者(有或无慢性肝病)进行了评估。我们比较了使用5次有效2D-SWE测量值的中位数或3次或5次有效2D-SWE测量值的均值时,TE评估的LS与2D-SWE测量值之间的相关性。

结果

我们在330/449例受试者(73.5%)中通过TE获得了可靠的LS测量值。其中,在281例受试者中我们获得了5次有效2D-SWE测量值。当我们使用5次有效2D-SWE测量值的中位数、5次有效2D-SWE测量值的均值或3次有效2D-SWE测量值的均值时,TE评估的LS与2D-SWE值之间的相关性相似:r分别为0.683、0.711和0.691。5次有效2D-SWE测量值的中位数、5次有效2D-SWE测量值的均值或3次有效2D-SWE测量值的均值之间无显著差异:分别为7.6 kPa、7.7 kPa和7.6 kPa。

结论

我们的研究表明,进行3次有效2D-SWE测量并使用这些测量值的均值就足够了。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验