Reliability of evoked endocardial T-wave sensing in 1,500 pacemaker patients.

作者信息

Boute W, Derrien Y, Wittkampf F H

出版信息

Pacing Clin Electrophysiol. 1986 Nov;9(6):948-53. doi: 10.1111/j.1540-8159.1986.tb06652.x.

Abstract

Intracardiac measurements of the evoked endocardial T-wave were performed on 1,500 pacemaker patients. The sensed evoked T-wave amplitude was found to be above 0.75 mV in 94.1 percent of all cases. No correlation was found between the amplitude of the spontaneous R-wave and that of the evoked T-wave. Some electrode characteristics proved to be more favorable than others for evoked endocardial wave sensing, i.e., a surface area of less than 12 mm2, porous surface structure (preferably carbon), and atraumatic fixation.