Eick Stefan, Loudermilk Brandon, Walberg Erik, Wente Moritz N
Aesculap AG, Am Aesculap-Platz, Tuttlingen, 78532, Germany.
Ann Surg Innov Res. 2013 Dec 10;7(1):15. doi: 10.1186/1750-1164-7-15.
In 1998, an electrothermal bipolar vessel sealing (EBVS) system was introduced and quickly became an integral component of the surgical armamentarium in various surgical specialties. Currently available EBVS instruments use a scissor-like jaw configuration and closing mechanism, which causes decreasing compression pressure from the proximal to the distal end of the jaws. A new EBVS system is described here which utilizes a different instrument jaw configuration and closing mechanism to enable a more homogeneous pressure distribution despite longer instrument jaws.
Results of jaw pressure distribution measurements as well as sealing experiments with subsequent burst pressure measurements ex vivo on bovine uterine arteries are demonstrated. Furthermore, an in vivo evaluation of the new EBVS system in a canine and porcine model including histological examination is presented.
The device revealed an even pressure distribution throughout the whole jaw length. The ex vivo burst pressure measurements revealed high average burst pressures, above 300 mmHg, independent of the outer diameter (1 to 7 mm) of the tested vessels. Histological evaluation of sealed vessels 21 days postoperatively demonstrated sealed and fused vessels without adjacent tissue damage.
The even pressure distribution leading to a sufficient tissue sealing in combination with the novel closing mechanism and extended jaw length differentiates the novel device from other available EBVS systems. This might offer a reduction of the overall procedure time, which should be further evaluated in a clinical study.
1998年,一种电热双极血管闭合(EBVS)系统被引入,并迅速成为各个外科专业手术器械库的一个组成部分。目前可用的EBVS器械采用类似剪刀的钳口配置和闭合机制,这导致钳口从近端到远端的压缩压力降低。本文描述了一种新的EBVS系统,该系统采用不同的器械钳口配置和闭合机制,尽管器械钳口更长,但仍能实现更均匀的压力分布。
展示了钳口压力分布测量结果以及随后在牛子宫动脉上进行的体外爆破压力测量的密封实验结果。此外,还介绍了在犬类和猪模型中对新EBVS系统进行的体内评估,包括组织学检查。
该装置在整个钳口长度上显示出均匀的压力分布。体外爆破压力测量显示平均爆破压力较高,超过300 mmHg,与测试血管的外径(1至7 mm)无关。术后21天对密封血管的组织学评估显示血管密封且融合,无相邻组织损伤。
均匀的压力分布导致足够的组织密封,再加上新颖的闭合机制和延长的钳口长度,使该新型装置与其他可用的EBVS系统有所不同。这可能会减少整个手术时间,应在临床研究中进一步评估。