具有自动自校准功能的多频电阻抗断层成像系统,用于长期监测。

Multi-frequency electrical impedance tomography system with automatic self-calibration for long-term monitoring.

作者信息

Wi Hun, Sohal Harsh, McEwan Alistair Lee, Woo Eung Je, Oh Tong In

出版信息

IEEE Trans Biomed Circuits Syst. 2014 Feb;8(1):119-28. doi: 10.1109/TBCAS.2013.2256785.

Abstract

Electrical Impedance Tomography (EIT) is a safe medical imaging technology, requiring no ionizing or heating radiation, as opposed to most other imaging modalities. This has led to a clinical interest in its use for long-term monitoring, possibly at the bedside, for ventilation monitoring, bleeding detection, gastric emptying and epilepsy foci diagnosis. These long-term applications demand auto-calibration and high stability over long time periods. To address this need we have developed a new multi-frequency EIT system called the KHU Mark2.5 with automatic self-calibration and cooperation with other devices via a timing signal for synchronization with other medical instruments. The impedance measurement module (IMM) for flexible configuration as a key component includes an independent constant current source, an independent differential voltmeter, and a current source calibrator, which allows automatic self-calibration of the current source within each IMM. We installed a resistor phantom inside the KHU Mark2.5 EIT system for intra-channel and inter-channel calibrations of all voltmeters in multiple IMMs. We show the deterioration of performance of an EIT system over time and the improvement due to automatic self-calibration. The system is able to maintain SNR of 80 dB for frequencies up to 250 kHz and below 0.5% reciprocity error over continuous operation for 24 hours. Automatic calibration at least every 3 days is shown to maintain SNR above 75 dB and reciprocity error below 0.7% over 7 days at 1 kHz. A clear degradation in performance results with increasing time between automatic calibrations allowing the tailoring of calibration to suit the performance requirements of each application.

摘要

电阻抗断层成像(EIT)是一种安全的医学成像技术,与大多数其他成像方式不同,它不需要电离辐射或热辐射。这使得临床上对其用于长期监测产生了兴趣,可能在床边进行,用于通气监测、出血检测、胃排空和癫痫病灶诊断。这些长期应用需要自动校准和长时间的高稳定性。为满足这一需求,我们开发了一种新的多频EIT系统,称为KHU Mark2.5,它具有自动自校准功能,并通过定时信号与其他设备协作,以便与其他医疗仪器同步。作为关键组件的用于灵活配置的阻抗测量模块(IMM)包括一个独立的恒流源、一个独立的差分电压表和一个电流源校准器,这允许在每个IMM内对电流源进行自动自校准。我们在KHU Mark2.5 EIT系统内部安装了一个电阻体模,用于对多个IMM中的所有电压表进行通道内和通道间校准。我们展示了EIT系统性能随时间的恶化以及自动自校准带来的改善。该系统在连续运行24小时的情况下,对于高达250 kHz的频率能够保持80 dB的信噪比,互易误差低于0.5%。结果表明,至少每3天进行一次自动校准,可在1 kHz频率下7天内保持信噪比高于75 dB,互易误差低于0.7%。随着自动校准之间时间间隔的增加,性能会明显下降,这使得可以根据每个应用的性能要求调整校准。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验