一种用于夜间无袖带血压测量的臂带式可穿戴设备。

An armband wearable device for overnight and cuff-less blood pressure measurement.

作者信息

Zheng Ya-Li, Yan Bryan P, Zhang Yuan-Ting, Poon Carmen C Y

出版信息

IEEE Trans Biomed Eng. 2014 Jul;61(7):2179-86. doi: 10.1109/TBME.2014.2318779. Epub 2014 Apr 18.

Abstract

24-h blood pressure (BP) has significant prognostic value for cardiovascular risk screening, but the present BP devices are mainly cuff-based, which are unsuitable for long-term BP measurement, especially during nighttime. In this paper, we developed an armband wearable pulse transit time (PTT) system for 24-h cuff-less BP measurement and evaluated it in an unattended out-of-laboratory setting. Ten healthy young subjects participated in this ambulatory study, where PTT was measured at 30-min interval by this wearable system and the reference BP was measured by a standard oscillometric ambulatory BP monitor. Due to the misalignment of BP and PTT on their recording time caused by the different measurement principles of the two BP devices, a new estimation method has been proposed: transients in PTT were removed from the raw data by defined criteria, and then evenly interpolated, low-pass filtered, and resampled to synchronize at the time when BP was recorded. The results show that with the proposed method, the correlation between PTT and systolic BP (SBP) during nighttime with dynamic range of 21.8 ± 14.2 mmHg has improved from -0.50 ± 0.24 to -0.62 ± 0.20 , and the difference between the estimated and reference SBP has improved from 0.7 ± 10.7 to 2.8 ± 8.2 mmHg with root mean square error reduced from 10.7 to 8.7 mmHg. In addition, the correlation between a very low frequency component of SBP and PTT obtained from the proposed method during nighttime is -0.80 ± 0.10 and the difference is 2.4 ± 5.7 mmHg for a dynamic BP range of 13.5 ± 8.0 mmHg. It is therefore concluded from this study that the proposed wearable system has great potential to be used for overnight SBP monitoring, especially to measure the averaged SBP over a long period.

摘要

24小时血压(BP)对心血管风险筛查具有重要的预后价值,但目前的血压测量设备主要基于袖带式,不适合长期血压测量,尤其是在夜间。在本文中,我们开发了一种用于24小时无袖带血压测量的臂带式可穿戴脉搏波传导时间(PTT)系统,并在无人值守的实验室外环境中对其进行了评估。十名健康的年轻受试者参与了这项动态研究,通过该可穿戴系统以30分钟的间隔测量PTT,并通过标准示波动态血压监测仪测量参考血压。由于两种血压测量设备的测量原理不同,导致血压和PTT在记录时间上存在偏差,因此提出了一种新的估计方法:通过定义的标准从原始数据中去除PTT中的瞬变,然后进行均匀插值、低通滤波和重采样,以使其在记录血压的时间同步。结果表明,采用该方法,夜间动态范围为21.8±14.2 mmHg时,PTT与收缩压(SBP)之间的相关性从-0.50±0.24提高到-0.62±0.20,估计SBP与参考SBP之间的差异从0.7±10.7 mmHg改善到2.8±8.2 mmHg,均方根误差从10.7 mmHg降低到8.7 mmHg。此外,夜间通过该方法获得的SBP极低频成分与PTT之间的相关性为-0.80±0.10,对于13.5±8.0 mmHg的动态血压范围,差异为2.4±5.7 mmHg。因此,本研究得出结论,所提出的可穿戴系统具有用于夜间SBP监测的巨大潜力,特别是用于长期测量平均SBP。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验