Alibek Sedat, Vogel Mika, Sun Wei, Winkler David, Baker Christopher A, Burke Michael, Gloger Hubertus
MVZ Radiologie and Nuklearmedizin, Fürth, Germany.
Diagn Interv Radiol. 2014 Jul-Aug;20(4):360-3. doi: 10.5152/dir.2014.13458.
Acoustic noise during magnetic resonance imaging (MRI) is the main source for patient discomfort and leads to verbal communication problems, difficulties in sedation, and hearing impairment. Silent Scan technology uses less changes in gradient excitation levels, which is directly related to noise levels. Here, we report our preliminary experience with this technique in neuroimaging with regard to subjective and objective noise levels and image quality.
Ten patients underwent routine brain MRI with 3 Tesla MR750w system and 12-channel head coil. T1-weighted gradient echo (BRAVO) and Silenz pulse sequence (TE=0, 3D radial center-out k-space filling and data sampling with relatively small gradient steps) were performed. Patients rated subjective sound impression for both sequences on a 6-point scale. Objective sound level measurements were performed with a dedicated device in gantry at different operation modes. Image quality was subjectively assessed in consensus by two radiologists on a 3-point scale.
Readers rated image quality as fully diagnostic in all patients. Measured mean noise was reduced significantly with Silenz sequence (68.8 dB vs. 104.65 dB with BRAVO, P = 0.024) corresponding to 34.3% reduction in sound intensity and 99,97% reduction in sound pressure. No significant difference was observed between Silenz sound levels and ambient sounds (i.e., background noise in the scanner room, 68.8 dB vs. 68.73 dB, P = 0.5). The patients' subjective sound level score was lower for Silenz compared with conventional sequence (1.1 vs. 2.3, P = 0.003).
T1-weighted Silent Scan is a promising technique for acoustic noise reduction and improved patient comfort.
磁共振成像(MRI)过程中的噪音是导致患者不适的主要原因,会引发言语交流问题、镇静困难及听力损害。静音扫描技术减少了梯度激发水平的变化,而梯度激发水平变化与噪音水平直接相关。在此,我们报告该技术在神经成像中关于主观和客观噪音水平及图像质量的初步经验。
10例患者使用3特斯拉MR750w系统和12通道头部线圈进行常规脑部MRI检查。分别采用T1加权梯度回波序列(BRAVO)和Silenz脉冲序列(TE = 0,3D径向中心向外k空间填充,采用相对较小的梯度步长进行数据采样)。患者对两种序列的主观声音感受按6分制进行评分。使用专门设备在不同操作模式下对机架内的客观声音水平进行测量。由两名放射科医生共同对图像质量进行主观评估,采用3分制。
阅片者认为所有患者的图像质量均可用于完全诊断。Silenz序列测得的平均噪音显著降低(68.8 dB,而BRAVO序列为104.65 dB,P = 0.024),对应声音强度降低34.3%,声压降低99.97%。Silenz序列的声音水平与环境声音之间未观察到显著差异(即扫描室内背景噪音,68.8 dB对68.73 dB,P = 0.5)。与传统序列相比,患者对Silenz序列的主观声音水平评分更低(1.1对2.3,P = 0.003)。
T1加权静音扫描是一种有前景的降低噪音及提高患者舒适度的技术。