To investigate the feasibility of 70 kV cerebral CT perfusion by comparing image quality and radiation exposure to 80 kV.
Thirty patients with suspected cerebral ischemia who underwent dual-source CT perfusion were divided into group A (80 kV, 150 mAs) and group B (70 kV, 150 mAs). Quantitative comparisons were used for maximum enhancement, signal-to-noise index (SNI), and values of cerebral blood flow (CBF), cerebral blood flow (CBV), mean transit time (MTT) on CBF, CBV, and MTT images, and radiation dose from these two groups. Qualitative perfusion images were assessed by two readers.
Maximum enhancement for group B was higher than group A (P < 0.05). There were no significant differences between the two groups for SNI on CBF and CBV maps (P = 0.06 - 0.576), but significant differences for MTT when SNI was measured on frontal white matter and temporo-occipital white matter (P < 0.05). There were no differences among values of CBF, CBV, and MTT for both groups (P = 0.251-0.917). Mean image quality score in group B was higher than group A for CBF (P < 0.05), but no differences for CBV (P = 0.542) and MTT (P = 0.962). Radiation dose for group B decreased compared with group A.
70 kV cerebral CT perfusion reduces radiation dose without compromising image quality.
• Radiation dose is a key concern with the increased using cerebral CT perfusion. Cerebral CT perfusion of 70 kV reduces radiation dose without compromising image quality. • A 70-kV protocol could be used for cerebral CT perfusion.
通过比较70 kV与80 kV脑CT灌注成像的图像质量及辐射剂量,探讨70 kV脑CT灌注成像的可行性。
30例疑似脑缺血患者接受双源CT灌注检查,分为A组(80 kV,150 mAs)和B组(70 kV,150 mAs)。对两组CBF、CBV和MTT图像上的最大强化值、信噪比指数(SNI)、脑血流量(CBF)、脑血容量(CBV)、平均通过时间(MTT)值以及辐射剂量进行定量比较。由两名阅片者对定性灌注图像进行评估。
B组的最大强化值高于A组(P < 0.05)。两组CBF和CBV图上的SNI无显著差异(P = 0.06 - 0.576),但在额叶白质和颞枕叶白质测量SNI时,MTT有显著差异(P < 0.05)。两组的CBF、CBV和MTT值无差异(P = 0.251 - 0.917)。B组CBF图像的平均图像质量评分高于A组(P < 0.05),但CBV(P = 0.542)和MTT(P = 0.962)无差异。B组的辐射剂量低于A组。
70 kV脑CT灌注成像可降低辐射剂量且不影响图像质量。
• 随着脑CT灌注成像应用的增加,辐射剂量成为关键问题。70 kV脑CT灌注成像可降低辐射剂量且不影响图像质量。• 70 kV方案可用于脑CT灌注成像。