The purpose of this study was to evaluate the feasibility of using a short echo time, three-dimensional H-1 magnetic resonance spectroscopic imaging (MRSI) sequence at 7 Tesla (T) to assess the metabolic signature of lesions for patients with glioma.
Twenty-nine patients with glioma were studied. MRSI data were obtained using CHESS water suppression, spectrally selective adiabatic inversion-recovery pulses and automatically prescribed outer-volume-suppression for lipid suppression, and spin echo slice selection (echo time = 30 ms). An interleaved flyback echo-planar trajectory was applied to shorten the total acquisition time (∼10 min). Relative metabolite ratios were estimated in tumor and in normal-appearing white and gray matter (NAWM, GM).
Levels of glutamine, myo-inositol, glycine, and glutathione relative to total creatine (tCr) were significantly increased in the T2 lesions for all tumor grades compared with those in the NAWM (P < 0.05), while N-acetyl aspartate to tCr were significantly decreased (P < 0.05). In grade 2 gliomas, level of total choline-containing-compounds to tCr was significantly increased (P = 0.0137), while glutamate to tCr was significantly reduced (P = 0.0012).
The improved sensitivity of MRSI and the increased number of metabolites that can be evaluated using 7T MR scanners is of interest for evaluating patients with glioma. This study has successfully demonstrated the application of a short-echo spin-echo MRSI sequence to detect characteristic differences in regions of tumor versus normal appearing brain.
本研究的目的是评估在7特斯拉(T)使用短回波时间的三维氢-1磁共振波谱成像(MRSI)序列评估胶质瘤患者病变代谢特征的可行性。
对29例胶质瘤患者进行了研究。使用CHESS水抑制、频谱选择性绝热反转恢复脉冲和自动设定的外容积抑制以抑制脂质,并采用自旋回波切片选择(回波时间 = 30 ms)获取MRSI数据。应用交错式回扫回波平面轨迹以缩短总采集时间(约10分钟)。估计肿瘤以及正常外观的白质和灰质(NAWM,GM)中的相对代谢物比率。
与NAWM相比,所有肿瘤分级的T2病变中谷氨酰胺、肌醇、甘氨酸和谷胱甘肽相对于总肌酸(tCr)的水平均显著升高(P < 0.05),而N-乙酰天门冬氨酸与tCr的比率则显著降低(P < 0.05)。在2级胶质瘤中,总含胆碱化合物与tCr的水平显著升高(P = 0.0137),而谷氨酸与tCr的水平则显著降低(P = 0.0012)。
MRSI灵敏度的提高以及使用7T MR扫描仪可评估的代谢物数量的增加,对于评估胶质瘤患者具有重要意义。本研究成功证明了短回波自旋回波MRSI序列在检测肿瘤区域与正常外观脑区特征差异方面的应用。