Almuhayawi Mohammed, Altun Osman, Strålin Kristoffer, Ozenci Volkan
Division of Clinical Microbiology, Karolinska Institutet, Karolinska University Hospital, Huddinge, Stockholm, Sweden King Abdul-Aziz University, Faculty of Medicine, Department of Microbiology, Jeddah, Saudi Arabia.
Division of Clinical Microbiology, Karolinska Institutet, Karolinska University Hospital, Huddinge, Stockholm, Sweden.
J Clin Microbiol. 2014 Sep;52(9):3230-6. doi: 10.1128/JCM.01084-14. Epub 2014 Jun 20.
In this study, we investigated the performance of the FilmArray and matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) in identifying microorganisms from blood culture (BC) bottles prior to positivity. First, we used simulated BacT/Alert FA Plus BC bottles with five each for Escherichia coli and Staphylococcus aureus isolates. The FilmArray identified all 10 isolates before BC positivity with 9/10 at 5 h and 1 at 7.5 h after incubation in the BC system. MALDI-TOF MS failed to identify the isolates prior to positivity. When the bottles were incubated for 2.5 h at room temperature (RT) before we put them into the BC system, the FilmArray identified 6/10 at 2.5 h and the remaining 4 at 5 h. Finally, we tested simulated BC bottles after incubation at RT. Interestingly, 9/10 isolates were identified with the FilmArray after 8 h of incubation at RT. Second, we studied clinical BC bottles in quadruplicate. When three-fourths of the parallel bottles signaled positive, the FilmArray was run on the fourth nonsignaled bottle and was found to be positive in 14/15 such cases. Third, we analyzed the performance of the FilmArray in the identification of microorganisms from clinical BC bottles before incubation in the system. Two milliliters of broth from 400 BC bottles was collected after arrival at the laboratory and stored at -70°C. Sixteen bottles later signaled positive in the system. When the frozen broth from these bottles was analyzed, the FilmArray identified all the microorganisms in 8/16 bottles prior to incubation in the BC system. This study shows that the FilmArray can identify microorganisms from BC bottles prior to positivity and in some cases even prior to incubation in the BC system.
在本研究中,我们调查了FilmArray和基质辅助激光解吸电离飞行时间质谱(MALDI-TOF MS)在血培养(BC)瓶阳性之前鉴定微生物的性能。首先,我们使用了模拟的BacT/Alert FA Plus BC瓶,每种分别装有5株大肠杆菌和金黄色葡萄球菌分离株。FilmArray在BC阳性之前鉴定出了所有10株分离株,在BC系统中孵育5小时时鉴定出9/10株,7.5小时时鉴定出1株。MALDI-TOF MS在阳性之前未能鉴定出分离株。当将瓶子在室温(RT)下孵育2.5小时后再放入BC系统时,FilmArray在2.5小时时鉴定出6/10株,其余4株在5小时时鉴定出。最后,我们测试了在RT下孵育后的模拟BC瓶。有趣的是,在RT下孵育8小时后,FilmArray鉴定出了9/10株分离株。其次,我们对临床BC瓶进行了一式四份的研究。当四分之三的平行瓶显示阳性时,对第四个未显示信号的瓶运行FilmArray,发现在14/15此类病例中呈阳性。第三,我们分析了FilmArray在系统孵育前从临床BC瓶中鉴定微生物的性能。在实验室收到400个BC瓶后,收集2毫升肉汤并储存在-70°C。后来有16个瓶在系统中显示阳性。当对这些瓶的冷冻肉汤进行分析时,FilmArray在BC系统孵育前在8/16个瓶中鉴定出了所有微生物。这项研究表明,FilmArray可以在阳性之前甚至在BC系统孵育之前从BC瓶中鉴定微生物。