Radiation dose verification of an X-ray based blood irradiator using EBT3 radiochromic films calibrated using Gamma Knife machine.

作者信息

Soliman Khaled, Adili Marouf, Alrushoud Abdullah

机构信息

Medical Physics Department, Prince Sultan Military Medical City, Riyadh, Saudi Arabia.

出版信息

Rep Pract Oncol Radiother. 2019 Jul-Aug;24(4):369-374. doi: 10.1016/j.rpor.2019.05.006. Epub 2019 Jun 7.

Abstract

AIM

Blood irradiators (BI) initial acceptance testing and routine annual dosimetry checks require radiation dose measurements in order to comply with regulatory requirements.

BACKGROUND

Traditionally thermo-luminescence dosimeters (TLD) have been used to measure the dose. The EBT3 film is reported to be a better dosimeter for low energy X-rays than its predecessors EBT2 and EBT. To the best of our knowledge, the use of EBT3 films to perform dosimetry on X-ray based BI has not been reported yet.

MATERIALS AND METHODS

We performed routine radiation dosimetry checks using EBT3 films on a new X-ray based BI and compared the results with TLD dosimetry. Calibration films were irradiated with radiation beam from a Co-60 Gamma Knife (GK) radiosurgery machine and, alternatively, using an Ir-192 high dose rate (HDR) brachytherapy device. The films were calibrated to cover a wide dose range from 1 to 40 Gy. Such a wide dose range has not been reported yet in BI film dosimetry.

RESULTS

We obtained a relative difference of about 6.6% between doses measured using TLD and those measured using EBT3 films. Both irradiation methods using GK or HDR were found to be adequate for the calibration of the EBT3 Gafchromic films.

CONCLUSIONS

We recommend the use of EBT3 films in routine X-ray based BI dosimetry checks. The presented method takes advantage of available radiotherapy equipment that can be efficiently used for EBT3 films calibration. The method is fast, reproducible and saves valuable medical physicist's time.

摘要

目的

血液辐照仪(BI)的初始验收测试和年度常规剂量测定检查需要进行辐射剂量测量,以符合监管要求。

背景

传统上一直使用热释光剂量计(TLD)来测量剂量。据报道,EBT3薄膜作为低能X射线剂量计比其前身EBT2和EBT性能更优。据我们所知,尚未有关于使用EBT3薄膜对基于X射线的BI进行剂量测定的报道。

材料与方法

我们使用EBT3薄膜对一台新型基于X射线的BI进行了常规辐射剂量测定检查,并将结果与TLD剂量测定结果进行了比较。校准薄膜分别用钴-60伽玛刀(GK)放射外科手术机器和铱-192高剂量率(HDR)近距离放射治疗设备的辐射束进行照射。这些薄膜经过校准,可覆盖1至40 Gy的宽剂量范围。在BI薄膜剂量测定中,如此宽的剂量范围尚未见报道。

结果

我们发现,使用TLD测量的剂量与使用EBT3薄膜测量的剂量之间的相对差异约为6.6%。结果表明,使用GK或HDR的两种照射方法均适用于EBT3变色薄膜的校准。

结论

我们建议在基于X射线的BI常规剂量测定检查中使用EBT3薄膜。所提出的方法利用了现有的放射治疗设备,可有效地用于EBT3薄膜校准。该方法快速、可重复,节省了医学物理师的宝贵时间。

相似文献

1
Radiation dose verification of an X-ray based blood irradiator using EBT3 radiochromic films calibrated using Gamma Knife machine.
Rep Pract Oncol Radiother. 2019 Jul-Aug;24(4):369-374. doi: 10.1016/j.rpor.2019.05.006. Epub 2019 Jun 7.
2
Quality Assurance of a 50-kV Radiotherapy Unit Using EBT3 GafChromic Film: A Feasibility Study.
Technol Cancer Res Treat. 2016 Feb;15(1):163-70. doi: 10.1177/1533034614565910. Epub 2015 Jan 9.
4
High dose-per-pulse electron beam dosimetry: Usability and dose-rate independence of EBT3 Gafchromic films.
Med Phys. 2017 Feb;44(2):725-735. doi: 10.1002/mp.12066. Epub 2017 Feb 8.
10
Radiochromic film dosimetry: considerations on precision and accuracy for EBT2 and EBT3 type films.
Z Med Phys. 2014 May;24(2):153-63. doi: 10.1016/j.zemedi.2013.08.002. Epub 2013 Sep 20.

引用本文的文献

2
Diacetylene-Based Colorimetric Radiation Sensors for the Detection and Measurement of γ Radiation during Blood Irradiation.
ACS Omega. 2021 Mar 29;6(14):9482-9491. doi: 10.1021/acsomega.0c06184. eCollection 2021 Apr 13.

本文引用的文献

1
Characteristic of EBT-XD and EBT3 radiochromic film dosimetry for photon and proton beams.
Phys Med Biol. 2018 Mar 15;63(6):065007. doi: 10.1088/1361-6560/aab1ee.
3
Gafchromic EBT-XD film: Dosimetry characterization in high-dose, volumetric-modulated arc therapy.
J Appl Clin Med Phys. 2016 Nov 8;17(6):312-322. doi: 10.1120/jacmp.v17i6.6281.
4
Evaluation of the uncertainty in an EBT3 film dosimetry system utilizing net optical density.
J Appl Clin Med Phys. 2016 Sep 8;17(5):466-481. doi: 10.1120/jacmp.v17i5.6262.
5
Gafchromic EBT3 film dosimetry in electron beams - energy dependence and improved film read-out.
J Appl Clin Med Phys. 2016 Jan 8;17(1):360-373. doi: 10.1120/jacmp.v17i1.5970.
6
Technical Note: On GAFChromic EBT-XD film and the lateral response artifact.
Med Phys. 2016 Feb;43(2):643-9. doi: 10.1118/1.4939226.
7
Technical Note: Initial characterization of the new EBT-XD Gafchromic film.
Med Phys. 2015 Oct;42(10):5782-6. doi: 10.1118/1.4930058.
9
Quality Assurance of a 50-kV Radiotherapy Unit Using EBT3 GafChromic Film: A Feasibility Study.
Technol Cancer Res Treat. 2016 Feb;15(1):163-70. doi: 10.1177/1533034614565910. Epub 2015 Jan 9.
10
Radiochromic film calibration for low-energy seed brachytherapy dose measurement.
Med Phys. 2014 Jul;41(7):072101. doi: 10.1118/1.4881146.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验