Suppr超能文献

使用碲锌镉单光子发射计算机断层扫描(CZT-SPECT)在心肌灌注成像中制定并验证个性化患者剂量方案。

Development and validation of a patient-tailored dose regime in myocardial perfusion imaging using CZT-SPECT.

作者信息

van Dijk J D, Jager P L, Mouden M, Slump C H, Ottervanger J P, de Boer J, Oostdijk A H J, van Dalen J A

机构信息

Department of Nuclear Medicine, Isala Hospital, Zwolle, The Netherlands,

出版信息

J Nucl Cardiol. 2014 Dec;21(6):1158-67. doi: 10.1007/s12350-014-9942-0. Epub 2014 Jul 9.

Abstract

BACKGROUND

Guidelines for SPECT myocardial perfusion imaging (MPI) traditionally recommend a fixed tracer dose. Yet, clinical practice shows degraded image quality in heavier patients. The aim was to optimize and validate the tracer dose and scan time to obtain a constant image quality less dependent on patients' physical characteristics.

METHODS

125 patients underwent Cadmium Zinc Telluride (CZT)-SPECT stress MPI using a fixed Tc-99m-tetrofosmin tracer dose. Image quality was scored by three physicians on a 4-point grading scale and related to the number of photon counts normalized to tracer dose and scan time. Counts were correlated with various patient-specific parameters dealing with patient size and weight to find the best predicting parameter. From these data, a formula to provide constant image quality was derived, and subsequently tested in 92 new patients.

RESULTS

Degradation in image quality and photon counts was observed for heavier patients for all patients' specific parameters (P < .01). We found body weight to be the best-predicting parameter for image quality and derived a new dose formula. After applying this new body weight-depended tracer dose and scan time in a new group, image quality was found to be constant (P > .19) in all patients.

CONCLUSIONS

Also in CZT SPECT image quality decreases with weight. The use of a tracer dose and scan time that depends linearly on patient's body weight corrected for the varying image quality in CZT-SPECT MPI. This leads to better radiation exposure justification.

摘要

背景

传统上,单光子发射计算机断层扫描心肌灌注成像(SPECT MPI)指南推荐使用固定的示踪剂剂量。然而,临床实践显示,体重较大的患者图像质量会下降。目的是优化并验证示踪剂剂量和扫描时间,以获得不那么依赖患者身体特征的恒定图像质量。

方法

125名患者使用固定剂量的锝-99m-替曲膦进行碲化镉锌(CZT)-SPECT负荷心肌灌注成像。由三名医生按照4分制对图像质量进行评分,并与根据示踪剂剂量和扫描时间标准化后的光子计数相关联。计数与各种与患者体型和体重相关的患者特异性参数相关,以找到最佳预测参数。根据这些数据,推导出一个提供恒定图像质量的公式,随后在92名新患者中进行测试。

结果

对于所有患者特异性参数,体重较大的患者图像质量和光子计数均出现下降(P <.01)。我们发现体重是图像质量的最佳预测参数,并推导出一个新的剂量公式。在新的一组患者中应用这个新的依赖体重的示踪剂剂量和扫描时间后,发现所有患者的图像质量都是恒定的(P >.19)。

结论

在CZT SPECT中,图像质量也会随着体重下降。在CZT-SPECT MPI中,使用与患者体重呈线性相关的示踪剂剂量和扫描时间,可校正变化的图像质量。这使得辐射暴露的合理性更高。