Suppr超能文献

CZT 相机中基于数据驱动的呼吸运动跟踪与补偿:对体模和人体图像的综合分析

Data-driven respiratory motion tracking and compensation in CZT cameras: a comprehensive analysis of phantom and human images.

作者信息

Ko Chi-Lun, Wu Yen-Wen, Cheng Mei-Fang, Yen Ruoh-Fang, Wu Wen-Chau, Tzen Kai-Yuan

机构信息

Department of Nuclear Medicine, National Taiwan University Hospital Yun-Lin Branch, No. 579, Sec. 2, Yunlin Rd., Douliou City, 640, Yunlin County, Taiwan,

出版信息

J Nucl Cardiol. 2015 Apr;22(2):308-18. doi: 10.1007/s12350-014-9963-8. Epub 2014 Aug 14.

Abstract

BACKGROUND

This study described a method for tracking and compensating respiratory motion in cadmium-zinc-telluride (CZT) cameras. We evaluated motion effects on myocardial perfusion imaging and assessed the usefulness of motion compensation in phantom and clinical studies.

METHODS

SPECT studies were obtained from an oscillating heart phantom and 552 patients using CZT cameras with list-mode acquisition. Images were reformatted in 500-ms frames, and the activity centroid was calculated as respiratory signal. The myocardial perfusion, left ventricular (LV) wall thickness, and LV volume were assessed before and after the motion compensation technique.

RESULTS

In phantom studies, we documented only minimal bias between simulated and measured shifts. Significantly reduced tracer activity, increased wall thickness and decreased volume in scans with 15 mm or more axial shifts were noted. In clinical studies, there was a higher prevalence of significant motion after treadmill exercise. The motion compensation technique could successfully compensate those motion artifacts.

CONCLUSION

The described method allows for tracking and compensating respiratory motion in CZT cameras. Significant respiratory motion is still not uncommon using CZT cameras, especially in patients who underwent treadmill tests. Motion blurring can be compensated using image processing techniques and image quality could be significantly improved.

摘要

背景

本研究描述了一种在碲锌镉(CZT)相机中跟踪和补偿呼吸运动的方法。我们评估了运动对心肌灌注成像的影响,并在体模和临床研究中评估了运动补偿的有效性。

方法

使用具有列表模式采集功能的CZT相机,从振荡心脏体模和552例患者中获取单光子发射计算机断层显像(SPECT)研究图像。图像以500毫秒的帧进行重新格式化,并将活动质心计算为呼吸信号。在运动补偿技术前后评估心肌灌注、左心室(LV)壁厚度和LV体积。

结果

在体模研究中,我们记录到模拟和测量的位移之间只有最小的偏差。在轴向位移为15毫米或更大的扫描中,发现示踪剂活性显著降低、壁厚度增加和体积减小。在临床研究中,跑步机运动后显著运动的发生率更高。运动补偿技术可以成功补偿这些运动伪影。

结论

所描述的方法允许在CZT相机中跟踪和补偿呼吸运动。使用CZT相机时,显著的呼吸运动仍然并不罕见,尤其是在接受跑步机测试的患者中。运动模糊可以通过图像处理技术进行补偿,并且图像质量可以显著提高。