Sannino Gianpaolo, Capparé Paolo, Gherlone Enrico F, Barlattani Alberto
Int J Oral Maxillofac Implants. 2015 Mar-Apr;30(2):351-8. doi: 10.11607/jomi.3747. Epub 2014 Dec 5.
The purpose of this study was to compare bone temperature changes during implant drilling with two drill designs employed in three different drilling sequences.
Two implant drill designs and three drilling sequences were evaluated in vitro using artificial bone cylinders. The evaluated drills were different only in the cutting-surface length (control, 16 mm; test, 4 mm). Three drilling sequences (control A, test B1, and test B2) were evaluated with and without irrigation. Temperatures were measured with thermocouple technology. The temperature changes generated by the final drill of each sequence were recorded as the experimental results and were subjected to the Student t test.
There were statistically significant differences in temperature changes when comparing the control group A with the test groups B1 (P = .001) and B2 (P = .01) during drilling without coolant. The mean temperature changes were 12.4°C, 6.5°C, and 13.7°C for groups A, B1, and B2, respectively. The Student t test showed statistically significant differences between temperature changes of the control group A and the test groups B1 (P < .01) and B2 (P < .05) during drilling with coolant. The mean temperature changes were 0.9°C, 0.7°C, and 1.9°C for groups A, B1, and B2, respectively.
Reduction in length of the cutting surface of the drill may limit frictional heat. Drills with the same length of cutting surface may induce lower bone temperature changes, when considering a preliminary drilling step with a pilot drill.
本研究的目的是比较在三种不同钻孔顺序下使用两种钻头设计进行种植体钻孔时的骨温度变化。
使用人工骨圆柱体在体外评估两种种植体钻头设计和三种钻孔顺序。所评估的钻头仅在切削面长度上有所不同(对照组,16毫米;试验组,4毫米)。对三种钻孔顺序(对照组A、试验组B1和试验组B2)在有和没有冲洗的情况下进行评估。使用热电偶技术测量温度。将每个顺序的最终钻头产生的温度变化记录为实验结果,并进行Student t检验。
在无冷却液钻孔过程中,将对照组A与试验组B1(P = 0.001)和B2(P = 0.01)比较时,温度变化存在统计学显著差异。A组、B1组和B2组的平均温度变化分别为12.4°C、6.5°C和13.7°C。Student t检验显示,在有冷却液钻孔过程中,对照组A与试验组B1(P < 0.01)和B2(P < 0.05)的温度变化之间存在统计学显著差异。A组、B1组和B2组的平均温度变化分别为0.9°C、0.7°C和1.9°C。
钻头切削面长度的减小可能会限制摩擦热。当考虑使用导向钻进行初步钻孔步骤时,具有相同切削面长度的钻头可能会引起较低的骨温度变化。