用于容积调强弧形治疗(VMAT)预处理剂量输出质量保证的ArcCHECK圆柱形二极管阵列的调试与综合评估。

Commissioning and comprehensive evaluation of the ArcCHECK cylindrical diode array for VMAT pretreatment delivery QA.

作者信息

Chaswal Vibha, Weldon Michael, Gupta Nilendu, Chakravarti Arnab, Rong Yi

机构信息

Ohio State University.

出版信息

J Appl Clin Med Phys. 2014 Jul 8;15(4):4832. doi: 10.1120/jacmp.v15i4.4832.

Abstract

We present commissioning and comprehensive evaluation for ArcCHECK as a QA equipment for volumetric-modulated arc therapy (VMAT), using the 6 MV photon beam with and without the flattening filter, and the SNC patient software (version 6.2). In addition to commissioning involving absolute dose calibration, array calibration, and PMMA density verification, ArcCHECK was evaluated for its response dependency on linac dose rate, instantaneous dose rate, radiation field size, beam angle, and couch insertion. Scatter dose characterization, consistency and symmetry of response, and dosimetry accuracy evaluation for fixed aperture arcs and clinical VMAT patient plans were also investigated. All the evaluation tests were performed with the central plug inserted and the homogeneous PMMA density value. Results of gamma analysis demonstrated an overall agreement between ArcCHECK-measured and TPS-calculated reference doses. The diode based field size dependency was found to be within 0.5% of the reference. The dose rate-based dependency was well within 1% of the TPS reference, and the angular dependency was found to be ± 3% of the reference, as tested for BEV angles, for both beams. Dosimetry of fixed arcs, using both narrow and wide field widths, resulted in clinically acceptable global gamma passing rates on the 3%/3mm level and 10% threshold. Dosimetry of narrow arcs showed an improvement over published literature. The clinical VMAT cases demonstrated high level of dosimetry accuracy in gamma passing rates.

摘要

我们展示了对ArcCHECK作为容积调强弧形放疗(VMAT)质量保证设备的调试和综合评估,使用了有无均整器的6 MV光子束以及SNC患者软件(版本6.2)。除了涉及绝对剂量校准、阵列校准和PMMA密度验证的调试外,还评估了ArcCHECK对直线加速器剂量率、瞬时剂量率、辐射野大小、射束角度和治疗床插入的响应依赖性。还研究了散射线剂量特性、响应的一致性和对称性,以及固定孔径弧形和临床VMAT患者计划的剂量测定准确性评估。所有评估测试均在插入中心插件且PMMA密度值均匀的情况下进行。伽马分析结果表明ArcCHECK测量的参考剂量与治疗计划系统(TPS)计算的参考剂量总体一致。发现基于二极管的野大小依赖性在参考值的0.5%以内。基于剂量率的依赖性远在TPS参考值的1%以内,并且对于两种射束,在BEV角度测试时,角度依赖性被发现为参考值的±3%。使用窄和宽野宽度的固定弧形的剂量测定在3%/3mm水平和10%阈值下产生了临床上可接受的总体伽马通过率。窄弧形的剂量测定显示比已发表的文献有所改进。临床VMAT病例在伽马通过率方面显示出高水平的剂量测定准确性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/c76d/5875521/39d41f6a1e17/ACM2-15-212-g001.jpg

相似文献

4
Initial clinical experience with ArcCHECK for IMRT/VMAT QA.
J Appl Clin Med Phys. 2016 Sep 8;17(5):20-33. doi: 10.1120/jacmp.v17i5.6118.
6
Verification of junction dose between VMAT arcs of total body irradiation using a Sun Nuclear ArcCHECK phantom.
J Appl Clin Med Phys. 2017 Nov;18(6):177-182. doi: 10.1002/acm2.12208. Epub 2017 Oct 29.
7
Quantitative evaluation of 3D dosimetry for stereotactic volumetric-modulated arc delivery using COMPASS.
J Appl Clin Med Phys. 2014 Jan 7;16(1):5128. doi: 10.1120/jacmp.v16i1.5128.
9
Rounded leaf end modeling in Pinnacle VMAT treatment planning for fixed jaw linacs.
J Appl Clin Med Phys. 2016 Nov 8;17(6):149-162. doi: 10.1120/jacmp.v17i6.6343.
10
Pixel response-based EPID dosimetry for patient specific QA.
J Appl Clin Med Phys. 2017 Jan;18(1):9-17. doi: 10.1002/acm2.12007. Epub 2016 Dec 15.

引用本文的文献

1
Characterization of a practically designed plastic scintillation plate dosimeter.
Med Phys. 2025 Jul;52(7):e17904. doi: 10.1002/mp.17904. Epub 2025 May 31.
2
Comparative Study of Fluence Distribution and Point Dose Using Arc-check and Delta Phantoms.
J Med Phys. 2024 Oct-Dec;49(4):706-709. doi: 10.4103/jmp.jmp_130_24. Epub 2024 Dec 18.
3
Assessing focal spot alignment in clinical linear accelerators: a comprehensive evaluation with triplet phantoms.
Phys Eng Sci Med. 2024 Dec;47(4):1361-1383. doi: 10.1007/s13246-024-01450-9. Epub 2024 Jul 2.
4
Technical note: A simple method for patient-specific quality assurance for lateral targets on a 1.5 T MR-Linac.
J Appl Clin Med Phys. 2024 Apr;25(4):e14323. doi: 10.1002/acm2.14323. Epub 2024 Mar 1.
5
Influence of dose calculation algorithms on the helical diode array using volumetric-modulated arc therapy for small targets.
J Appl Clin Med Phys. 2024 Jul;25(7):e14307. doi: 10.1002/acm2.14307. Epub 2024 Feb 16.
6
Comparison of Progressive Resolution Optimizer and Photon Optimizer algorithms in RapidArc delivery for head and neck SIB treatments.
Rep Pract Oncol Radiother. 2023 Nov 16;28(5):623-635. doi: 10.5603/rpor.97431. eCollection 2023.
7
Commissioning and clinical evaluation of a novel high-resolution quality assurance digital detector array for SRS and SBRT.
J Appl Clin Med Phys. 2024 Apr;25(4):e14258. doi: 10.1002/acm2.14258. Epub 2024 Jan 4.
8
Preliminary study on dosimetry characteristics of a novel cylindrical dose verification system.
J Appl Clin Med Phys. 2023 Oct;24(10):e14138. doi: 10.1002/acm2.14138. Epub 2023 Sep 4.
9
10
Assessment of Delivery Quality Assurance for Stereotactic Radiosurgery With Cyberknife.
Front Oncol. 2021 Nov 17;11:751922. doi: 10.3389/fonc.2021.751922. eCollection 2021.

本文引用的文献

2
Delivery quality assurance with ArcCHECK.
Med Dosim. 2013 Spring;38(1):77-80. doi: 10.1016/j.meddos.2012.07.004. Epub 2012 Aug 21.
3
Evaluation of the ArcCHECK QA system for IMRT and VMAT verification.
Phys Med. 2013 May;29(3):295-303. doi: 10.1016/j.ejmp.2012.04.005. Epub 2012 May 12.
4
5
Evaluation of a new VMAT QA device, or the "X" and "O" array geometries.
J Appl Clin Med Phys. 2011 Jan 31;12(2):3346. doi: 10.1120/jacmp.v12i2.3346.
6
The accuracy of dose calculations by anisotropic analytical algorithms for stereotactic radiotherapy in nasopharyngeal carcinoma.
Phys Med Biol. 2011 Jan 21;56(2):397-413. doi: 10.1088/0031-9155/56/2/008. Epub 2010 Dec 22.
7
Commissioning of photon beams of a flattening filter-free linear accelerator and the accuracy of beam modeling using an anisotropic analytical algorithm.
Int J Radiat Oncol Biol Phys. 2011 Jul 15;80(4):1228-37. doi: 10.1016/j.ijrobp.2010.09.050. Epub 2010 Dec 2.
8
Accuracy of out-of-field dose calculations by a commercial treatment planning system.
Phys Med Biol. 2010 Dec 7;55(23):6999-7008. doi: 10.1088/0031-9155/55/23/S03. Epub 2010 Nov 12.
10

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验