Chaswal Vibha, Weldon Michael, Gupta Nilendu, Chakravarti Arnab, Rong Yi
Ohio State University.
J Appl Clin Med Phys. 2014 Jul 8;15(4):4832. doi: 10.1120/jacmp.v15i4.4832.
We present commissioning and comprehensive evaluation for ArcCHECK as a QA equipment for volumetric-modulated arc therapy (VMAT), using the 6 MV photon beam with and without the flattening filter, and the SNC patient software (version 6.2). In addition to commissioning involving absolute dose calibration, array calibration, and PMMA density verification, ArcCHECK was evaluated for its response dependency on linac dose rate, instantaneous dose rate, radiation field size, beam angle, and couch insertion. Scatter dose characterization, consistency and symmetry of response, and dosimetry accuracy evaluation for fixed aperture arcs and clinical VMAT patient plans were also investigated. All the evaluation tests were performed with the central plug inserted and the homogeneous PMMA density value. Results of gamma analysis demonstrated an overall agreement between ArcCHECK-measured and TPS-calculated reference doses. The diode based field size dependency was found to be within 0.5% of the reference. The dose rate-based dependency was well within 1% of the TPS reference, and the angular dependency was found to be ± 3% of the reference, as tested for BEV angles, for both beams. Dosimetry of fixed arcs, using both narrow and wide field widths, resulted in clinically acceptable global gamma passing rates on the 3%/3mm level and 10% threshold. Dosimetry of narrow arcs showed an improvement over published literature. The clinical VMAT cases demonstrated high level of dosimetry accuracy in gamma passing rates.
我们展示了对ArcCHECK作为容积调强弧形放疗(VMAT)质量保证设备的调试和综合评估,使用了有无均整器的6 MV光子束以及SNC患者软件(版本6.2)。除了涉及绝对剂量校准、阵列校准和PMMA密度验证的调试外,还评估了ArcCHECK对直线加速器剂量率、瞬时剂量率、辐射野大小、射束角度和治疗床插入的响应依赖性。还研究了散射线剂量特性、响应的一致性和对称性,以及固定孔径弧形和临床VMAT患者计划的剂量测定准确性评估。所有评估测试均在插入中心插件且PMMA密度值均匀的情况下进行。伽马分析结果表明ArcCHECK测量的参考剂量与治疗计划系统(TPS)计算的参考剂量总体一致。发现基于二极管的野大小依赖性在参考值的0.5%以内。基于剂量率的依赖性远在TPS参考值的1%以内,并且对于两种射束,在BEV角度测试时,角度依赖性被发现为参考值的±3%。使用窄和宽野宽度的固定弧形的剂量测定在3%/3mm水平和10%阈值下产生了临床上可接受的总体伽马通过率。窄弧形的剂量测定显示比已发表的文献有所改进。临床VMAT病例在伽马通过率方面显示出高水平的剂量测定准确性。