Suppr超能文献

人类供体角膜和病理角膜的全视野光学相干断层扫描

Full-field optical coherence tomography of human donor and pathological corneas.

作者信息

Ghouali Wajdene, Grieve Kate, Bellefqih Salima, Sandali Otman, Harms Fabrice, Laroche Laurent, Paques Michel, Borderie Vincent

机构信息

Institut de la Vision, UPMC Univ Paris 06 , UMR_S 968, Paris, F-75012 , France .

出版信息

Curr Eye Res. 2015 May;40(5):526-34. doi: 10.3109/02713683.2014.935444. Epub 2014 Sep 24.

Abstract

PURPOSE

To evaluate the performance of a full-field optical coherence tomography (FF-OCT) system in the study of human donor and pathological corneas and assess its suitability for use in eye banks.

METHODS

Our study was carried out using an FF-OCT system developed for non-invasive imaging of tissue structures in depth with ultrahigh resolution (1 µm in all directions). Images were acquired from eight stored human donor corneas (either edematous or after deswelling) and five surgical specimens of corneas with various diseases (bullous keratopathy, lattice corneal dystrophy, stromal scar after keratitis, keratoconus and Fuchs dystrophy). They were compared with standard histology and pre-operative spectral domain OCT.

RESULTS

The FF-OCT device enabled a precise visualization of the cells and the different structures (epithelium, basement membrane, Bowman's layer, stroma, Descemet's membrane and endothelium) in normal corneas. Specific lesions in various corneal diseases could also be easily identified, such as corneal edema, epithelium and Bowman's layer irregularities, breaks, or scars (keratoconus), stromal opacities, deposits, fibrosis (stromal corneal scar, bullous keratopathy, lattice corneal dystrophy) and Descemet's membrane thickening and guttae (Fuchs dystrophy). FF-OCT image features were comparable to the details provided by conventional histology. Higher resolution could be demonstrated with FF-OCT when compared with spectral domain OCT.

CONCLUSION

FF-OCT is a powerful non-invasive imaging tool that allows detailed study of corneal structures. Images correlate well with conventional histology. Further studies should evaluate the benefit of this technique as a complement to current assessment methods of human donor corneas.

摘要

目的

评估全视野光学相干断层扫描(FF - OCT)系统在研究人类供体角膜和病理角膜方面的性能,并评估其在眼库中的适用性。

方法

我们使用为深度非侵入性成像组织结构而开发的超高分辨率(各方向均为1 µm)的FF - OCT系统进行研究。从八个储存的人类供体角膜(水肿或消肿后)和五个患有各种疾病(大疱性角膜病变、格子状角膜营养不良、角膜炎后基质瘢痕、圆锥角膜和富克斯角膜营养不良)的角膜手术标本中获取图像。将它们与标准组织学和术前光谱域OCT进行比较。

结果

FF - OCT设备能够精确可视化正常角膜中的细胞和不同结构(上皮、基底膜、Bowman层、基质、Descemet膜和内皮)。各种角膜疾病中的特定病变也很容易识别,如角膜水肿、上皮和Bowman层不规则、断裂或瘢痕(圆锥角膜)、基质混浊、沉积物、纤维化(基质角膜瘢痕、大疱性角膜病变、格子状角膜营养不良)以及Descemet膜增厚和滴状赘生物(富克斯角膜营养不良)。FF - OCT图像特征与传统组织学提供的细节相当。与光谱域OCT相比,FF - OCT可显示更高的分辨率。

结论

FF - OCT是一种强大的非侵入性成像工具,可对角膜结构进行详细研究。图像与传统组织学相关性良好。进一步的研究应评估该技术作为人类供体角膜当前评估方法补充的益处。