仅在近附加度数上有所不同的两种多焦点人工晶状体设计的比较。

Comparison of two multifocal intraocular lens designs that differ only in near add.

作者信息

Pedrotti Emilio, Mastropasqua Rodolfo, Passilongo Mattia, Parisi Graziella, Marchesoni Ivan, Marchini Giorgio

出版信息

J Refract Surg. 2014 Nov;30(11):754-60. doi: 10.3928/1081597X-20141021-07.

Abstract

PURPOSE

To evaluate monocular functional outcomes after the implantation of the AcrySof ReSTOR SN6AD2 intraocular lens (IOL) (+2.5 diopters [D] near add) (Alcon Laboratories, Inc., Fort Worth, TX) and the AcrySof ReSTOR SN6AD1 IOL (+3.0 D near add) (Alcon Laboratories, Inc.).

METHODS

This prospective, comparative, nonrandomized single-blind observational study comprised 62 eyes of 62 patients that underwent phacoemulsification and implantation of a multifocal IOL: SN6AD2 (31 eyes) (+2.5 group) and SN6AD1 (31 eyes) (+3.0 group). Twelve months after surgery, monocular near (30 and 40 cm), intermediate (50, 60, and 70 cm), and distance (4 m) visual acuity were evaluated with the internal root mean square and modulation transfer function. Both parameters were evaluated at the 4- and 6-mm pupil sizes.

RESULTS

No statistical differences at 4 m were found between the groups. The +2.5 group obtained better performances at all intermediate distances (50 cm, 0.23 ± 0.14 vs 0.32 ± 0.13; 60 and 70 cm, 0.21 ± 0.10 vs 0.41 ± 0.14 and 0.24 ± 0.10 vs 0.53 ± 0.17, respectively), whereas the near visual acuity was better for the +3.0 group (30 and 40 cm, 0.36 ± 0.18 vs 0.14 ± 0.09 and 0.36 ± 0.19 vs 0.17 ± 0.07, respectively). The root mean square was lower for the +2.5 group compared to the +3.0 group, whereas the modulation transfer function showed overlapping results between the two models.

CONCLUSIONS

Both IOL models showed good results in distance vision; the +2.5 D IOL seemed to provide better intermediate vision than the +3.0 D IOL. For near vision, the +3.0 D model performed better than the +2.5 D model.

摘要

目的

评估植入爱尔康公司(位于德克萨斯州沃思堡)的AcrySof ReSTOR SN6AD2人工晶状体(IOL)(近附加度数为+2.5屈光度[D])和AcrySof ReSTOR SN6AD1 IOL(近附加度数为+3.0 D)后的单眼功能结果。

方法

这项前瞻性、比较性、非随机单盲观察性研究纳入了62例患者的62只眼,这些患者接受了白内障超声乳化吸除术并植入了多焦点IOL:SN6AD2(31只眼)(+2.5组)和SN6AD1(31只眼)(+3.0组)。术后12个月,使用内均方根和调制传递函数评估单眼近视力(30和40 cm)、中视力(50、60和70 cm)及远视力(4 m)。在4 mm和6 mm瞳孔直径下对这两个参数进行评估。

结果

两组在4 m处未发现统计学差异。+2.5组在所有中距离(50 cm,0.23±0.14对0.32±0.13;60和70 cm,分别为0.21±0.10对0.41±0.14以及0.24±0.10对0.53±0.17)表现更佳,而+3.0组的近视力更好(30和40 cm,分别为0.36±0.18对0.14±0.09以及0.36±0.19对0.17±0.07)。+2.5组的均方根低于+3.0组,而调制传递函数显示两种型号之间的结果有重叠。

结论

两种IOL型号在远视力方面均显示出良好结果;+2.5 D的IOL似乎比+3.0 D的IOL提供更好的中视力。对于近视力,+3.0 D型号比+2.5 D型号表现更好。