眼部眼轴长度的光学生物测量与压平式超声测量的比较。

Comparison of optical biometry and applanation ultrasound measurements of the axial length of the eye.

作者信息

Nakhli Fouad R

机构信息

Optometry and Investigation Services, Dhahran Eye Specialist Hospital (DESH), P.O. Box 39455 Zip: 31942, Dhahran, Eastern Province, Saudi Arabia.

出版信息

Saudi J Ophthalmol. 2014 Oct;28(4):287-91. doi: 10.1016/j.sjopt.2014.04.003. Epub 2014 May 6.

Abstract

PURPOSE

To investigate the relationship between optical biometry and applanation ultrasound measurement of the axial length of the eye.

MATERIALS AND METHODS

This prospective study enrolled 55 (68 eyes) consecutive patients scheduled for cataract surgery at Dhahran Eye Specialist Hospital (DESH). Every eye underwent two measurements each with an optical biometer and with applanation ultrasound. Only patients with cataract and no other ophthalmic pathology or disease were enrolled. Comparison, correlation and repeatability of axial length with both devices were analyzed. Agreement between devices was evaluated. A regression formula to convert measurements between devices was investigated. A p value less than 0.05 was statistically significant.

RESULTS

There was strong repeatability (99.4%) and agreement (r = 0.987) between both devices (p < 0.001); the difference between devices was mainly in short eyes (p = 0.031).

CONCLUSION

Optical biometry and applanation ultrasound measurements of axial length correlate well. However, optical biometry is preferable in short eyes.

摘要

目的

研究光学生物测量与压平式超声测量眼轴长度之间的关系。

材料与方法

这项前瞻性研究纳入了达兰眼科专科医院(DESH)计划进行白内障手术的55例(68只眼)连续患者。每只眼睛分别使用光学生物测量仪和压平式超声进行两次测量。仅纳入患有白内障且无其他眼科病理或疾病的患者。分析了两种设备测量眼轴长度的比较、相关性和可重复性。评估了两种设备之间的一致性。研究了用于在两种设备之间转换测量值的回归公式。p值小于0.05具有统计学意义。

结果

两种设备之间具有很强的可重复性(99.4%)和一致性(r = 0.987)(p < 0.001);两种设备之间的差异主要存在于短眼(p = 0.031)。

结论

光学生物测量和压平式超声测量眼轴长度具有良好的相关性。然而,在短眼中光学生物测量更可取。

相似文献

1
Comparison of optical biometry and applanation ultrasound measurements of the axial length of the eye.
Saudi J Ophthalmol. 2014 Oct;28(4):287-91. doi: 10.1016/j.sjopt.2014.04.003. Epub 2014 May 6.
3
Comparison of biometric measurements using partial coherence interferometry and applanation ultrasound.
J Cataract Refract Surg. 2003 Apr;29(4):747-52. doi: 10.1016/s0886-3350(02)01739-x.
5
Comparison of ocular biometry and intraocular lens power using a new biometer and a standard biometer.
J Cataract Refract Surg. 2014 May;40(5):709-15. doi: 10.1016/j.jcrs.2013.09.020. Epub 2014 Mar 18.
6
Comparison of the Zeiss IOLMaster and applanation A-scan ultrasound: biometry for intraocular lens calculation.
Clin Exp Ophthalmol. 2003 Apr;31(2):121-4. doi: 10.1046/j.1442-9071.2003.00617.x.

引用本文的文献

2
Comparison of optical biometry versus ultrasound biometry in different axial length groups.
Saudi J Ophthalmol. 2024 Mar 1;39(2):198-205. doi: 10.4103/sjopt.sjopt_267_23. eCollection 2025 Apr-Jun.
3
Evaluation of Axial Length and Refractive Outcomes in Patients With Dense Vitreous Hemorrhage Who Have Phacovitrectomy.
J Vitreoretin Dis. 2025 Feb 19:24741264251322211. doi: 10.1177/24741264251322211.
4
Volume measurement of the vitrectomised eye and its applications in practice.
Indian J Ophthalmol. 2024 Oct 1;72(10):1501-1505. doi: 10.4103/IJO.IJO_3237_23. Epub 2024 Sep 27.
5
Anatomic Characteristics of Eyes With Fuchs Endothelial Corneal Dystrophy.
Cornea. 2025 Mar 1;44(3):311-323. doi: 10.1097/ICO.0000000000003694. Epub 2024 Sep 4.
6
Comparison of the ocular ultrasonic and optical biometry devices in the different quality measurements.
J Optom. 2023 Oct-Dec;16(4):284-295. doi: 10.1016/j.optom.2023.05.001. Epub 2023 Aug 9.
8
Refractive Outcomes after Cataract Surgery.
Diagnostics (Basel). 2022 Jan 19;12(2):243. doi: 10.3390/diagnostics12020243.
9
Biometry characteristics in congenital cataract patients before surgery in a tertiary eye care centre in Nepal.
Saudi J Ophthalmol. 2019 Oct-Dec;33(4):342-346. doi: 10.1016/j.sjopt.2019.07.004. Epub 2019 Jul 17.
10
Monochromatic and white light and the regulation of eye growth.
Exp Eye Res. 2019 Jul;184:172-182. doi: 10.1016/j.exer.2019.04.020. Epub 2019 Apr 21.

本文引用的文献

2
Comparison of Optical versus Ultrasonic Biometry in Keratoconic Eyes.
J Ophthalmol. 2013;2013:481238. doi: 10.1155/2013/481238. Epub 2013 Aug 5.
3
Biometric measurements in highly myopic eyes.
J Cataract Refract Surg. 2013 Feb;39(2):180-7. doi: 10.1016/j.jcrs.2012.08.064. Epub 2012 Dec 7.
4
Comparison of optical low-coherence reflectometry and applanation ultrasound biometry on intraocular lens power calculation.
Graefes Arch Clin Exp Ophthalmol. 2011 Jan;249(1):69-75. doi: 10.1007/s00417-010-1509-4. Epub 2010 Sep 18.
5
Comparison of 2 laser instruments for measuring axial length.
J Cataract Refract Surg. 2010 Apr;36(4):644-8. doi: 10.1016/j.jcrs.2009.11.007.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验