肝脏磁共振弹性成像的跨厂商验证

Cross-vendor validation of liver magnetic resonance elastography.

作者信息

Serai Suraj D, Yin Meng, Wang Hui, Ehman Richard L, Podberesky Daniel J

机构信息

Department of Radiology, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA,

出版信息

Abdom Imaging. 2015 Apr;40(4):789-94. doi: 10.1007/s00261-014-0282-y.

Abstract

PURPOSE

To evaluate and validate the reproducibility of MR Elastography (MRE)-derived liver stiffness values on two different MR vendor platforms performed on the same subject on the same day.

METHODS

This investigation was approved by the hospital IRB. MRE exams were performed twice in identical fashion in eight volunteers and in five clinical patients on two different 1.5 T MR scanners-once on a Philips MR scanner and immediately afterward in back-to-back fashion on a General Electric MR scanner, or vice versa. All scan parameters were kept identical on the two platforms to the best extent possible. After the MRE magnitude and phase images were obtained, the data were converted into quantitative images displaying the stiffness of the liver parenchyma. Mean liver stiffness values between the two platforms were compared using interclass correlation with a p value <0.05 considered statistically significant.

RESULTS

Interclass correlation coefficient (ICC) value of 0.994 was obtained for 13 subjects with p value <0.001 indicating a significantly positive correlation.

CONCLUSION

As MRE gains in acceptance and as its availability becomes more widespread, it is important to ascertain and confirm that liver stiffness values obtained on different MRE vendor platforms are consistent and reproducible. In this small pilot investigation, we demonstrate that liver stiffness measurement with MRE is reproducible and has very good consistency across two vendor platforms.

摘要

目的

评估并验证同一天在同一受试者身上使用两种不同磁共振成像(MRI)供应商平台进行磁共振弹性成像(MRE)得出的肝脏硬度值的可重复性。

方法

本研究经医院机构审查委员会批准。8名志愿者和5名临床患者在两台不同的1.5T MR扫描仪上以相同方式进行了两次MRE检查——一次在飞利浦MR扫描仪上进行,随后立即以背靠背方式在通用电气MR扫描仪上进行,反之亦然。两个平台上的所有扫描参数尽可能保持相同。在获得MRE幅度和相位图像后,将数据转换为显示肝实质硬度的定量图像。使用组内相关性比较两个平台之间的平均肝脏硬度值,p值<0.05被认为具有统计学意义。

结果

13名受试者的组内相关系数(ICC)值为0.994,p值<0.001,表明存在显著正相关。

结论

随着MRE的认可度提高及其可用性更加广泛,确定并确认在不同MRE供应商平台上获得的肝脏硬度值具有一致性和可重复性非常重要。在这项小型初步研究中,我们证明了使用MRE进行肝脏硬度测量具有可重复性,并且在两个供应商平台之间具有非常好的一致性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/c2c1/4523216/81af6a372e2b/nihms708787f1.jpg

相似文献

引用本文的文献

10
Liver fibrosis assessment: MR and US elastography.肝脏纤维化评估:磁共振和超声弹性成像。
Abdom Radiol (NY). 2022 Sep;47(9):3037-3050. doi: 10.1007/s00261-021-03269-4. Epub 2021 Oct 23.

本文引用的文献

8
Pediatric liver MR elastography.小儿肝脏磁共振弹性成像。
Dig Dis Sci. 2012 Oct;57(10):2713-9. doi: 10.1007/s10620-012-2196-2. Epub 2012 May 9.
9
Agreed statistics: measurement method comparison.一致的统计数据:测量方法比较。
Anesthesiology. 2012 Jan;116(1):182-5. doi: 10.1097/ALN.0b013e31823d7784.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验