Wahlers Thorsten C, Rahmanian Parwis B
Department of Cardiothoracic Surgery, Heart Center University Hospital Cologne, Cologne, Germany.
Thorac Cardiovasc Surg. 2015 Sep;63(6):504-7. doi: 10.1055/s-0035-1546298. Epub 2015 Mar 10.
In aortic valve replacement, severe calcification of the sinotubular junction is a limitation in the application of the rapid deployment Edwards Intuity valve system. This is illustrated by a case presentation in which discrepancies of shape and diameter between sizer and valve mounted on the deployment system resulted in discarding the prosthesis and performing a biocomposite root replacement instead. Modification of the valve sizer will allow for safer sizing because the sizing process will mimic the implantation process more accurately.