The objective of the present study was to compare QUANTA Flash dsDNA, a chemiluminescent immunoassay (CIA) on the BIO-FLASH, a rapid-response chemiluminescent analyzer, to three other anti-dsDNA antibody assays and to Crithidia luciliae indirect immunofluorescence test (CLIFT).
In the first part of the study, 161 samples, 61 from patients suffering from systemic lupus erythematosus (SLE) and 100 from a disease control group, were tested by QUANTA Flash dsDNA CIA, QUANTA Lite dsDNA SC ELISA, BioPlex 2200 multiplex flow immunoassay (MFI), ImmuLisa dsDNA ELISA, and NOVA Lite CLIFT. A second cohort of 69 SLE patients was then tested by QUANTA Flash dsDNA and CLIFT to expand the study.
The overall qualitative agreements varied between 77.0% (NOVA Lite CLIFT versus QUANTA Lite) and 89.4% (ImmuLisa versus NOVA Lite CLIFT). The clinical sensitivities for the anti-dsDNA antibody tests varied from 8.2% (NOVA Lite CLIFT) to 54.1% (QUANTA Lite), while the clinical specificities varied from 88.0% (BioPlex 2200) to 100.0% (NOVA Lite CLIFT). Good correlation was found between QUANTA Flash dsDNA and NOVA Lite CLIFT.
Significant variations among dsDNA methods were observed. QUANTA Flash dsDNA provides a good combination of sensitivity and specificity for the diagnosis of SLE and good agreement to CLIFT.
本研究的目的是比较 QUANTA Flash dsDNA,一种基于 BIO-FLASH 的化学发光免疫分析(CIA),与另外三种抗 dsDNA 抗体检测方法和克氏锥虫间接免疫荧光试验(CLIFT)。
在研究的第一部分,对 161 份样本进行了检测,其中 61 份来自系统性红斑狼疮(SLE)患者,100 份来自疾病对照组。检测方法包括 QUANTA Flash dsDNA CIA、QUANTA Lite dsDNA SC ELISA、BioPlex 2200 多重流式免疫分析(MFI)、ImmuLisa dsDNA ELISA 和 NOVA Lite CLIFT。然后,对另一组 69 例 SLE 患者进行了 QUANTA Flash dsDNA 和 CLIFT 检测,以扩大研究范围。
整体定性一致性在 77.0%(NOVA Lite CLIFT 与 QUANTA Lite)到 89.4%(ImmuLisa 与 NOVA Lite CLIFT)之间。抗 dsDNA 抗体检测的临床敏感性从 8.2%(NOVA Lite CLIFT)到 54.1%(QUANTA Lite)不等,而临床特异性从 88.0%(BioPlex 2200)到 100.0%(NOVA Lite CLIFT)不等。QUANTA Flash dsDNA 与 NOVA Lite CLIFT 之间存在良好的相关性。
dsDNA 方法之间存在显著差异。QUANTA Flash dsDNA 为 SLE 的诊断提供了良好的敏感性和特异性组合,与 CLIFT 具有良好的一致性。