使用两种类型的微视野计设备进行固视稳定性测量。

Fixation Stability Measurement Using Two Types of Microperimetry Devices.

作者信息

Liu Hongting, Bittencourt Millena G, Sophie Raafay, Sepah Yasir J, Hanout Mostafa, Rentiya Zubir, Annam Rachel, Scholl Hendrik P N, Nguyen Quan Dong

机构信息

Wilmer Eye Institute, Johns Hopkins University, Baltimore, Maryland, USA ; Visual Science and Optometry Center, People's Hospital of Guangxi Zhuang Autonomous Region, Nanning, Guangxi Province, China.

Wilmer Eye Institute, Johns Hopkins University, Baltimore, Maryland, USA.

出版信息

Transl Vis Sci Technol. 2015 Mar 10;4(2):3. doi: 10.1167/tvst.4.2.3. eCollection 2015 Mar.

Abstract

PURPOSE

We compared the fixation stability measurements obtained with two microperimeters, the Micro Perimeter 1 (MP-1) and the Spectral OCT/SLO (OCT/SLO), in subjects with and without maculopathies.

METHODS

A total of 41 eyes with no known ocular diseases and 45 eyes with maculopathies were enrolled in the study. Both eyes of each participant had a 20-second fixation test using the MP-1 and OCT/SLO. The bivariate contour ellipse area (BCEA) was used for fixation stability evaluation.

RESULTS

In the normal group, BCEA was 2.93 ± 0.32 log minarc on OCT/SLO and 2.89 ± 0.30 log minarc on MP-1. In the maculopathy group, BCEA was 3.05 ± 0.41 log minarc on OCT/SLO and 3.15 ± 0.46 log minarc on MP-1. There was no statistically significant difference between the BCEA measured by OCT/SLO and by MP-1 in both groups. A moderate correlation was found between the two devices ( = 0.45, < 0.001). The sample size during the fixation test was 535.5 ± 14.6 pairs of coordinates in the normal group and 530.7 ± 14.9 pairs in the maculopathy group with MP-1, while it was 72.3 ± 6.9 and 59.9 ± 10.1, respectively, with OCT/SLO. This was due to different tracking frequencies between the two microperimeters.

CONCLUSION

Fixation stability assessment yields similar results using the OCT/SLO and MP-1. A major difference in sampling rate between the two microperimeters does not significantly affect BCEA measurements.

TRANSLATIONAL RELEVANCE

Fixation stability assessments are comparable and interchangeable between the OCT/SLO and the MP-1.

摘要

目的

我们比较了在患有和未患有黄斑病变的受试者中,使用两种微视野计(微视野计1(MP-1)和光谱光学相干断层扫描/扫描激光检眼镜(OCT/SLO))获得的注视稳定性测量结果。

方法

共有41只无已知眼部疾病的眼睛和45只患有黄斑病变的眼睛纳入本研究。每位参与者的双眼均使用MP-1和OCT/SLO进行了20秒的注视测试。使用双变量轮廓椭圆面积(BCEA)评估注视稳定性。

结果

在正常组中,OCT/SLO上的BCEA为2.93±0.32对数分钟弧度,MP-1上为2.89±0.30对数分钟弧度。在黄斑病变组中,OCT/SLO上的BCEA为3.05±0.41对数分钟弧度,MP-1上为3.15±0.46对数分钟弧度。两组中OCT/SLO和MP-1测量的BCEA之间无统计学显著差异。两种设备之间发现中度相关性(=0.45,<0.001)。使用MP-1时,正常组注视测试期间的样本量为535.5±14.6对坐标,黄斑病变组为530.7±14.9对;而使用OCT/SLO时,分别为72.3±6.9和59.9±10.1。这是由于两种微视野计的跟踪频率不同。

结论

使用OCT/SLO和MP-1进行注视稳定性评估产生相似的结果。两种微视野计采样率的主要差异不会显著影响BCEA测量。

转化相关性

OCT/SLO和MP-1之间的注视稳定性评估具有可比性且可互换。

相似文献

1
Fixation Stability Measurement Using Two Types of Microperimetry Devices.
Transl Vis Sci Technol. 2015 Mar 10;4(2):3. doi: 10.1167/tvst.4.2.3. eCollection 2015 Mar.
2
Assessment of Central Retinal Sensitivity Employing Two Types of Microperimetry Devices.
Transl Vis Sci Technol. 2014 Sep 12;3(5):3. doi: 10.1167/tvst.3.5.3. eCollection 2014 Sep.
4
A microperimeter that provides fixation pattern and retinal sensitivity measurement.
Jpn J Ophthalmol. 2006 Mar-Apr;50(2):111-5. doi: 10.1007/s10384-005-0292-y.
5
Alterations in Fixation Indices in Primary Open-Angle Glaucoma by Microperimetry.
J Clin Med. 2022 Apr 23;11(9):2368. doi: 10.3390/jcm11092368.
6
Static and dynamic retinal fixation stability in microperimetry.
Can J Ophthalmol. 2013 Oct;48(5):375-80. doi: 10.1016/j.jcjo.2013.05.021.
7
Efficacy of SLO-Microperimetry and Humphrey for evaluating macular sensitivity changes in advanced glaucoma.
Can J Ophthalmol. 2013 Oct;48(5):406-12. doi: 10.1016/j.jcjo.2013.08.001.
8
Reference Clinical Database for Fixation Stability Metrics in Normal Subjects Measured with the MAIA Microperimeter.
Transl Vis Sci Technol. 2016 Nov 15;5(6):6. doi: 10.1167/tvst.5.6.6. eCollection 2016 Nov.

引用本文的文献

1
3
Inter and intradevice assessment of microperimetry testing in aging eyes.
Sci Rep. 2024 Jan 10;14(1):1049. doi: 10.1038/s41598-024-51539-0.
4
Visual function and biofeedback training of patients with central vision loss: a review.
Int J Ophthalmol. 2023 May 18;16(5):824-831. doi: 10.18240/ijo.2023.05.21. eCollection 2023.
6
Evaluation of Fixational Behavior throughout Life.
Brain Sci. 2021 Dec 24;12(1):19. doi: 10.3390/brainsci12010019.
7
Detecting Diabetic Retinal Neuropathy Using Fundus Perimetry.
Int J Mol Sci. 2021 Oct 3;22(19):10726. doi: 10.3390/ijms221910726.
8
Metrics and Acquisition Modes for Fixation Stability as a Visual Function Biomarker.
Invest Ophthalmol Vis Sci. 2017 May 1;58(6):BIO268-BIO276. doi: 10.1167/iovs.17-21710.
9
Reference Clinical Database for Fixation Stability Metrics in Normal Subjects Measured with the MAIA Microperimeter.
Transl Vis Sci Technol. 2016 Nov 15;5(6):6. doi: 10.1167/tvst.5.6.6. eCollection 2016 Nov.

本文引用的文献

1
Assessment of Central Retinal Sensitivity Employing Two Types of Microperimetry Devices.
Transl Vis Sci Technol. 2014 Sep 12;3(5):3. doi: 10.1167/tvst.3.5.3. eCollection 2014 Sep.
3
Normal values for fundus perimetry with the microperimeter MP1.
Ophthalmology. 2010 Aug;117(8):1571-6, 1576.e1. doi: 10.1016/j.ophtha.2009.12.044. Epub 2010 May 15.
5
Fixation stability measurement using the MP1 microperimeter.
Retina. 2009 May;29(5):651-6. doi: 10.1097/IAE.0b013e318196bd65.
6
Retinal fixation impairment in diabetic macular edema.
Retina. 2008 Nov-Dec;28(10):1443-50. doi: 10.1097/IAE.0b013e318183571e.
8
Fundus microperimetry patterns of fixation in type 2 diabetic patients with diffuse macular edema.
Retina. 2007 Jan;27(1):21-9. doi: 10.1097/01.iae.0000256658.71864.ca.
9
Fixation stability using radial gratings in patients with age-related macular degeneration.
Can J Ophthalmol. 2006 Jun;41(3):333-9. doi: 10.1139/I06-019.
10
Combined use of SLO microperimetry and OCT for retinal functional and structural testing.
Graefes Arch Clin Exp Ophthalmol. 2006 May;244(5):634-8. doi: 10.1007/s00417-005-0088-2. Epub 2005 Aug 25.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验