Department of Ophthalmology and Visual Sciences, University of Illinois at Chicago, IL 60612, USA.
Eye (Lond). 2011 Feb;25(2):245-51. doi: 10.1038/eye.2010.158. Epub 2010 Oct 29.
To establish normative values for macular light sensitivity and to determine the intrasession fluctuation of perimetric responses using the OPKO/OTI microperimeter.
A total of 32 visually normal subjects participated in the study. A standardized grid pattern was used for testing, which consisted of 28 points arranged concentrically in three circles that occupied an area of 11° (in diameter) within the central macula. Each subject participated in at least two tests. Parameters evaluated included: overall mean macular sensitivity for test 1 and 2, overall difference in mean macular sensitivity between tests, and the mean sensitivity for each circle. The relationship between sensitivity and age was also examined.
The overall median sensitivity for test 1 was 16.8 decibels (dB) and for test 2 was 16.9 dB. The median sensitivities for test 1 and test 2 were not significantly different (P = 0.72). The mean intrasession sensitivity difference was 0.13 dB. The variability of the sensitivity difference between tests decreased as mean sensitivity increased. The sensitivity values averaged across the two tests for inner, middle, and outer circles ranged from 14.3 to 18.8 dB (median value of 16.9 dB), 13.8-18.3 dB (median value of 17.2 dB), and 11.3-18.3 dB (median value of 16.6 dB), respectively. Linear regression analysis showed a 0.5 dB sensitivity loss for each decade of life.
We documented a narrow range of intrasession fluctuation using the OPKO/OTI microperimeter. The establishment of normative sensitivity values will facilitate monitoring the loss of macular visual function in patients with retinal disease.
利用 OPKO/OTI 微视野计建立黄斑光敏感度的正常值,并确定视野计检查的日内波动。
共有 32 名视力正常的受试者参与了这项研究。使用标准化的网格图案进行测试,该图案由 28 个点组成,以 3 个同心圆排列,占据中央黄斑内 11°(直径)的区域。每个受试者至少参加了两次测试。评估的参数包括:第 1 次和第 2 次测试的平均黄斑总体敏感性、两次测试的平均黄斑敏感性差异以及每个圆的平均敏感性。还检查了敏感性与年龄之间的关系。
第 1 次测试的总体中位数敏感性为 16.8 分贝(dB),第 2 次测试为 16.9 dB。第 1 次和第 2 次测试的中位数敏感性无显著差异(P = 0.72)。平均日内敏感性差异为 0.13 dB。随着平均敏感性的增加,测试间敏感性差异的变异性减小。两次测试的平均敏感性值在内部、中部和外部圆圈中的范围为 14.3 至 18.8 dB(中位数为 16.9 dB)、13.8-18.3 dB(中位数为 17.2 dB)和 11.3-18.3 dB(中位数为 16.6 dB)。线性回归分析显示,每 10 年生命会损失 0.5 dB 的敏感性。
我们使用 OPKO/OTI 微视野计记录了一个狭窄的日内波动范围。建立正常的敏感性值将有助于监测视网膜疾病患者黄斑视觉功能的丧失。