Factors affecting intradiscal pressure measurement during in vitro biomechanical tests.

作者信息

Tremblay Jaëlle, Brailovski Vladimir, Mac-Thiong Jean-Marc, Petit Yvan

机构信息

Department of Mechanical Engineering, École de technologie supérieure, Montreal, H3C 1K3, Canada ; Research center, Hôpital du Sacré-Cœur, Montreal, H4J 1C5, Canada.

Research center, Hôpital du Sacré-Cœur, Montreal, H4J 1C5, Canada.

出版信息

Scoliosis. 2015 Feb 11;10(Suppl 2):S1. doi: 10.1186/1748-7161-10-S2-S1. eCollection 2015.

Abstract

OBJECTIVES

To assess the reliability of intradiscal pressure measurement during in vitro biomechanical testing. In particular, the variability of measurements will be assessed for repeated measures by considering the effect of specimens and of freezing/thawing cycles.

METHODS

Thirty-six functional units from 8 porcine spines (S1: T7-T8, S2: T9-T10, S3: T12-T11, S4: T14-T13, S5: L1-L2 and S6: L3-L4) have been used. The intervertebral discs were measured to obtain the frontal and sagittal dimensions. These measurements helped locate the center of the disc where a modified catheter was positioned. A fiber optic pressure sensor (measuring range: -0.1 to 17 bar) (360HP, SAMBA Sensors, Sweden) was then inserted into the catheter. The specimens were divided into 3 groups: 1) fresh (F), 2) after one freeze/thaw cycle (C1) and 3) after 2 freeze/thaw cycles (C2). These groups were divided in two, depending on whether specimens were subjected to 400 N axial loading or not. Ten measurements (insertion of the sensor for a period of one minute, then removal) were taken for each case. Statistical analyses evaluated the influence of porcine specimen and the vertebral level using a MANOVA. The effect of repeated measurements was evaluated with ANOVA. The difference between freeze/thaw cycles were analysed with U Mann-Whitney test (P≤0.05).

RESULTS

Without axial loading, the F group showed 365 mbar intradiscal pressure, 473 mbar for the C1 group, and 391 mbar for the C2 group. With 400N axial load, the F group showed intradiscal pressure of 10610 mbar, the C1 group 10132 mbar, the C2 group 12074 mbar. The statistical analysis shows a significant influence of the porcine specimen (p<0.001), with or without axial loading and of the vertebral level with (p=0.048) and without load (p<0.001). The results were also significantly different between the freeze/thaw cycles, with (p<0.001) and without load (p=0.033). Repeated measurement (without load p = 0.82 and with p = 0.56) did not show significant influence.

CONCLUSIONS

The results tend to support that freezing/thawing cycles can affect intradiscal pressure measurement with significant inter-specimen variability. The use of the same specimen as its own control during in vitro biomechanical testing could be recommended.

摘要

目的

评估体外生物力学测试期间椎间盘内压力测量的可靠性。特别是,将通过考虑标本以及冻融循环的影响来评估重复测量的测量变异性。

方法

使用了来自8头猪脊柱的36个功能单元(S1:T7 - T8,S2:T9 - T10,S3:T12 - T11,S4:T14 - T13,S5:L1 - L2和S6:L3 - L4)。测量椎间盘以获得前后径和矢状径尺寸。这些测量有助于确定椎间盘中心位置,在此处放置改良导管。然后将光纤压力传感器(测量范围:-0.1至17巴)(360HP,SAMBA传感器,瑞典)插入导管。标本分为3组:1)新鲜组(F),2)经过1次冻融循环后(C1)和3)经过2次冻融循环后(C2)。根据标本是否承受400N轴向载荷,将这些组再分为两组。每种情况进行10次测量(传感器插入1分钟,然后取出)。统计分析使用多变量方差分析评估猪标本和椎体水平的影响。使用方差分析评估重复测量的影响。使用曼 - 惠特尼U检验分析冻融循环之间的差异(P≤0.05)。

结果

无轴向载荷时,F组椎间盘内压力显示为365毫巴,C1组为473毫巴,C2组为391毫巴。施加400N轴向载荷时,F组椎间盘内压力显示为10610毫巴,C1组为10132毫巴,C2组为12074毫巴。统计分析表明,无论有无轴向载荷,猪标本均有显著影响(p < 0.001),椎体水平在有载荷时(p = 0.048)和无载荷时(p < 0.001)也有显著影响。冻融循环之间的结果在有载荷时(p < 0.001)和无载荷时(p = 0.033)也有显著差异。重复测量(无载荷时p = 0.82,有载荷时p = 0.56)未显示显著影响。

结论

结果倾向于支持冻融循环会影响椎间盘内压力测量,且标本间存在显著变异性。建议在体外生物力学测试期间使用同一标本作为自身对照。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/1b0b/4331726/7fb490af10f2/1748-7161-10-S2-S1-1.jpg

相似文献

7
In vitro disc pressure profiles below scoliosis fusion constructs.脊柱侧弯融合结构下方的体外椎间盘压力分布
Spine (Phila Pa 1976). 2008 Sep 15;33(20):2134-42. doi: 10.1097/BRS.0b013e31817d1d7f.