Nakamura Masakazu, Iso Hiroyasu, Kitamura Akihiko, Imano Hironori, Kiyama Masahiko, Yokoyama Shinji, Kayamori Yuzo, Koyama Isao, Nishimura Kunihiro, Nakai Michikazu, Dasti Mahnaz, Vesper Hubert W, Teramoto Tamio, Miyamoto Yoshihiro
National Cerebral and Cardiovascular Center, Department of Preventive Cardiology, Lipid Reference Laboratory, Japan.
Public Health, Department of Social Medicine, Graduate School of Medicine, Osaka University, Japan.
Clin Chim Acta. 2015 May 20;445:127-32. doi: 10.1016/j.cca.2015.03.026. Epub 2015 Mar 26.
Accurate measurement of total cholesterol (TC) is important for cardiovascular disease risk management. The US Centers for Disease Control and Prevention (CDC) and Cholesterol Reference Method Laboratory Network (CRMLN) perform Abell-Levy-Brodie-Kendall (AK) reference measurement procedure (RMP) for TC as a secondary reference method, and implement Certification Protocol for Manufacturers. Japanese CRMLN laboratory at Osaka performed the AK RMP for 22 years, and conducted TC certification for reagent/calibrator/instrument systems of six Japanese manufacturers every 2 years for 16 years. Osaka TC performance was examined and compared to CDC's reference values.
AK RMP involved sample hydrolysis, cholesterol extraction, and determination of cholesterol levels by spectrophotometry. The Certification Protocol for Manufacturers includes comparison with AK RMP using at least 40 fresh specimens. Demonstration of average bias ≤3% and total coefficient of variation ≤3% qualified an analytical system for certification.
In the AK RMP used in the Osaka CRMLN laboratory, the regression equation for measuring TC was y (Osaka)=1.000x (CDC)+0.032 (n=619, R(2)=1.000). Six Japanese manufacturers had allowable performance for certification.
The AK RMP for TC measurement was accurate, precise, and stable for 22 years. Six Japanese manufacturers were certified for 16 years.
准确测量总胆固醇(TC)对于心血管疾病风险管理至关重要。美国疾病控制与预防中心(CDC)和胆固醇参考方法实验室网络(CRMLN)采用阿贝尔-利维-布罗迪-肯德尔(AK)参考测量程序(RMP)作为TC的二级参考方法,并实施制造商认证协议。大阪的日本CRMLN实验室进行了22年的AK RMP,并在16年里每两年对6家日本制造商的试剂/校准品/仪器系统进行TC认证。对大阪的TC性能进行了检测,并与CDC的参考值进行了比较。
AK RMP包括样品水解、胆固醇提取以及通过分光光度法测定胆固醇水平。制造商认证协议包括使用至少40个新鲜标本与AK RMP进行比较。平均偏差≤3%且总变异系数≤3%证明分析系统符合认证要求。
在大阪CRMLN实验室使用的AK RMP中,测量TC的回归方程为y(大阪)=1.000x(CDC)+0.032(n = 619,R² = 1.000)。6家日本制造商的性能符合认证要求。
用于TC测量的AK RMP在22年里准确、精密且稳定。6家日本制造商在16年里获得了认证。