Treviño-Garza Consuelo, Mancillas-Adame Leonardo, Estrada-Zúñiga Cynthia M, Villarreal-Pérez Jesús Z, Villarreal-Martinez Laura, De la O-Cavazos Manuel E
Department of Pediatrics, Autonomous University of Nuevo Leon, Monterrey, Mexico.
Division of Endocrinology, Autonomous University of Nuevo Leon, Monterrey, Mexico.
J Clin Lab Anal. 2016 May;30(3):254-7. doi: 10.1002/jcla.21845. Epub 2015 Mar 26.
Assays based on multiplex immunoassay (MIA) technology have demonstrated advantages over enzyme-linked immunosorbent assay (ELISA) and radioimmunoassay (RIA). Its acceptance depends on how well it performs in comparison to older techniques. The aim is to compare the results of leptin using RIA versus MIA.
We analyzed 81 samples of umbilical cord blood of healthy term newborns by RIA and MIA.
The concordance correlation coefficient was 0.158 (95% CI 0.10-0.21). Pearson's correlation coefficient was 0.6651 (95% CI 0.52-0.77; P < 0.0001). In the Bland-Altman plot, concordance is acceptable because most of the measurements are within a mean of ±1.96 SD.
As shown by the Bland-Altman plot, there is concordance by both methods, but with a weak correlation.
基于多重免疫分析(MIA)技术的检测方法已显示出优于酶联免疫吸附测定(ELISA)和放射免疫测定(RIA)的优势。其认可度取决于与旧技术相比的性能表现。目的是比较使用RIA和MIA检测瘦素的结果。
我们采用RIA和MIA分析了81例健康足月儿的脐带血样本。
一致性相关系数为0.158(95%可信区间0.10 - 0.21)。Pearson相关系数为0.6651(95%可信区间0.52 - 0.77;P < 0.0001)。在Bland - Altman图中,一致性是可接受的,因为大多数测量值在均值±1.96标准差范围内。
如Bland - Altman图所示,两种方法存在一致性,但相关性较弱。