Cirrus HD-OCT视网膜追踪系统在眼球运动期间视网膜神经纤维层厚度测量的可重复性

Retinal Nerve Fiber Layer Thickness Measurement Repeatability for Cirrus HD-OCT Retinal Tracking System During Eye Movement.

作者信息

Hwang Young Hoon, Song Miryoung, Kim Dai Woo, Uhm Ki Bang

机构信息

*Department of Ophthalmology, Kim's Eye Hospital, Myung-Gok Eye Research Institute, Konyang University †Department of Ophthalmology, Sanggye Paik Hospital, Inje University College of Medicine ‡Department of Ophthalmology, Hanyang University College of Medicine, Seoul, Korea.

出版信息

J Glaucoma. 2016 Mar;25(3):e214-9. doi: 10.1097/IJG.0000000000000250.

Abstract

PURPOSE

To investigate the repeatability of peripapillary retinal nerve fiber layer (RNFL) thickness measurements obtained using Cirrus high-definition optical coherence tomographic (Cirrus HD-OCT) retinal tracking system during various types of eye movements.

MATERIALS AND METHODS

We included 20 healthy eyes, 40 glaucomatous eyes of elderly patients, and 17 eyes with pathologic nystagmus. For healthy eyes, RNFL thickness measurements were obtained under 3 conditions: (1) without eye movement, fixated on the device's internal target, (2) with horizontal eye movement, and (3) with vertical eye movement during scan acquisition. Each session was performed 3 times with and without the use of the retinal tracking system. The repeatability of RNFL thickness measurements obtained with and without the retinal tracking was compared within each session and among the sessions.

RESULTS

In healthy eyes, measurements obtained without the use of a retinal tracking system showed lower repeatability when measurements were obtained with eye movements than without (P<0.05). However, when retinal tracking system was used, measurement repeatability under conditions with eye movements was similar to that achieved without eye movement (P>0.05) and higher than that obtained without the use of a retinal tracking system (P<0.05). In elderly glaucomatous patients, the retinal tacking system did not significantly improve measurement repeatability during fixation (P>0.05). In eyes with pathologic nystagmus, the tracking system did not follow eye movement and scan acquisition was not processed.

CONCLUSIONS

Cirrus HD-OCT retinal tracking system may enhance RNFL thickness measurement repeatability under certain, but not all, eye movement conditions.

摘要

目的

研究使用Cirrus高清光学相干断层扫描(Cirrus HD-OCT)视网膜追踪系统在各种眼球运动类型下获得的视乳头周围视网膜神经纤维层(RNFL)厚度测量的可重复性。

材料与方法

我们纳入了20只健康眼睛、40只老年患者的青光眼眼睛以及17只患有病理性眼球震颤的眼睛。对于健康眼睛,在3种条件下获得RNFL厚度测量值:(1)无眼球运动,注视设备内部目标;(2)水平眼球运动;(3)扫描采集期间垂直眼球运动。每次测量在使用和不使用视网膜追踪系统的情况下各进行3次。比较每次测量中使用和不使用视网膜追踪获得的RNFL厚度测量的可重复性,以及各次测量之间的可重复性。

结果

在健康眼睛中,不使用视网膜追踪系统进行测量时,眼球运动时获得的测量值的可重复性低于无眼球运动时(P<0.05)。然而,当使用视网膜追踪系统时,眼球运动条件下的测量可重复性与无眼球运动时相似(P>0.05),且高于不使用视网膜追踪系统时获得的可重复性(P<0.05)。在老年青光眼患者中,视网膜追踪系统在注视期间未显著提高测量可重复性(P>0.05)。在患有病理性眼球震颤的眼睛中,追踪系统无法跟踪眼球运动,扫描采集未进行。

结论

Cirrus HD-OCT视网膜追踪系统可能在某些但并非所有眼球运动条件下提高RNFL厚度测量的可重复性。